Identification Details and Hardware Changelog


Nominal Operating Voltages and Currents

Ch.LineVI





Performance Measures

Experiment=detdaq18, Run=222 (Date)

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

FH3268 ± 2314829
FM3205 ± 2254777
FL3175 ± 2214758
AHL3271 ± 2294842
AML3207 ± 2244797

Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (sigma)Number of NCP
FH6419276
FM3393224
FL2402881
AHL7133023
AML3383411

     

Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH1
FM 1
FL 6017
AHL157
AML 71

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History


ExperimentHutchShiftsDate Deployed
CommissioningXCS115 July 2020