Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram

FH3303 ± 2334847Graphs
FM3232 ± 2264804Graphs
FL3199 ± 2214790Graphs
AHL3306 ± 2314858Graphs
AML3234 ± 2254822Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain Mode

Variation (ADU)

Number of NCP

Pixels Map, Histogram

FH6461801Graphs
FM3378589Graphs
FL2358870Graphs
AHL7166346Graphs
AML3376682     Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain Mode

Number of NCP

FH1
FM1
FL6017
AHL171
AML80

Configuration Settings:

 

Linearity (signal injection)

Thermal measurements (all thermometers)


Experiment History

ExperimentHutchShiftsDate Deployed
Commissioning