Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram

FH3297 ± 2324846Graphs
FM3238 ± 2264811Graphs
FL3207 ± 2214808Graphs
AHL3304 ± 2304856Graphs
AML3240 ± 2244828Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain Mode

Variation (ADU)

Number of NCP

Pixels Map, Histogram

FH61491Graphs
FM34360Graphs
FL21976Graphs
AHL74865Graphs
AML32150     Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain Mode

Number of NCP

FH1
FM1
FL5589
AHL144
AML67

Configuration Settings:

  

Linearity (signal injection)

Thermal measurements (all thermometers)


Experiment History

ExperimentHutchShiftsDate Deployed
Commissioning