Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram

FH3268 ± 2314829Graphs
FM3205 ± 2254777Graphs
FL3175 ± 2214758Graphs
AHL3271 ± 2294842Graphs
AML3207 ± 2244797Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain Mode

Variation (ADU)

Number of NCP

Pixels Map, Histogram

FH6419276Graphs
FM3393224Graphs
FL2402881Graphs
AHL7133023Graphs
AML3383411     Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain Mode

Number of NCP

FH1
FM 1
FL 6017
AHL157
AML 71

Configuration Settings:

 

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History

ExperimentHutchShiftsDate Deployed
Commissioning