Half-module # | Contains hybrid | Contains sensor | Hybrid glued | Sensor glued | Sent to UCSC | HV wirebonded | Signals wirebonded | QA tested | Results | Sent to SLAC | Mounted to cooling block | Part of module | Notes |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
to fixture | lousy | with a prayer | mashed bonds | painfully | ? test 1 | 4 chips bonded, and 4 are alive. 3 of them are both bonded and alive. | ? | never |
| ||||
2/23/12(TKN) | 2/23/12(TKN) | 2/23/12(AG) | 2/27/12(FM) | 2/29/12(FM) | OK: test 2 | 3/12/12(TKN) | ? | READY | |||||
3/1/12(TKN) | 3/1/12(TKN) | 3/1/12(OM) | 3/2/12(FM) | 3/6/12(FM) | 3/5/12 test 1 | OK | 3/14/12(PH) | ? | Damaged signal wirebonds observed: probable shorts first ~40 channels. | ||||
3/10/12(TKN) | 3/10/12(TKN) | 3/12/12 (TKN) | 3/12/12(FM) | 3/13/12(FM) | 3/13/12 test 1 | High noise? (To retest) Chan 625 unbonded, 627 dead. |
|
|
| Cont'ed on next line | |||
Cont'ed No 3 |
|
|
|
|
|
| Chan 625 fixed, 3/22/12(FM) | 3/23/12 test 6 | 627 dead, else OK: test 6 | 3/23/12 (PH) | ? | Needs ITT connector | |
3/10/12(TKN) | 3/10/12(TKN) | 3/12/12 (TKN) | 3/14/12(FM) | 3/14/12(FM) | Channels 317, 319 shorted, rest is OK: test 1 | 3/18/12 (VF) | ? | READY | |||||
3/10/12(TKN) | 3/10/12(TKN) | 3/12/12 (TKN) | 3/14/12(FM) | 3/14/12(FM) | 3/19/12 (VF,OM) test 1 | OK: test 5 (but cal7_8x3_125ns is bad for this run) | 3/23/12(PH) | ? | Needs ITT connector | ||||
3/18/12(TKN) | 3/18/12(TKN) | 3/19/12 (TKN/VF) | 3/21/12(FM) noticed scratch on Bias/Guard rings. But IV is OK up to 300V. | 3/22/12(FM) | OK except channels 0 (no response), 134, 639 (noisy response). Channel 0 is bad b/c of the scratch that connects it with bias ring. | 3/23/12(PH) | ? | READY | |||||
3/18/12(TKN) | 3/18/12(TKN) | 3/19/12 (TKN/VF) | 3/21/12(FM) | 3/22/12(FM) | 3/23/12 test 1 | OK: test 11 | 3/27/12(OM) | ? | READY | ||||
3/18/12(TKN) | 3/18/12(TKN) | 3/19/12 (TKN/VF) | 3/21/12(FM) | 3/21/12(FM) | OK: test 3 | 3/27/12(OM) | ? | READY | |||||
3/20/12(TKN) | 3/20/12(TKN) | 3/21/12(DG) | 3/23/12(FM) | 3/23/12(FM) | 3/26/12 test 1 | Channel 466 has no response (as on hybrid), else OK. | 3/27/12(OM) | ? | READY | ||||
3/20/12(TKN) | 3/20/12(TKN) | 3/21/12(DG) | 3/23/12(FM) | 3/23/12(FM) | 3/26/12 test 1 | Channels 81 and 83 shorted? Else OK. | 3/27/12(PH) | ? | READY | ||||
3/20/12(TKN) | 3/20/12(TKN) | 3/21/12(DG) | 3/23/12(FM) | 3/23/12(FM) | 3/26/12 test 1 | Shorts between channels 262 and 264 as well as between channels 403 and 405; channel 183 dead and high noise on channel 184; else OK. | 3/27/12(PH) | ? | READY | ||||
3/22/12(TKN) | 3/22/12(TKN) | 3/23/12(PH) | 03/27/12(FM) | 03/27/12(FM) | 03/27/12(OM)test 1 test 2 | Channel 345 dead (as on hybrid), else OK: test 31 | 3/31/12(OM) | ? | READY | ||||
3/22/12(TKN) | 3/22/12(TKN) | 3/23/12(PH) | 03/28/12(FM) | 03/28/12(FM) | 03/29/12(OM)test 1 | OK: test 5 | 3/31/12(OM) | ? | READY | ||||
3/22/12(TKN) | 3/22/12(TKN) | 3/23/12(PH) | 03/28/12(FM) | 03/28/12(FM) | 3/29/12 (OM) test 1 | Pedestal shift on channel 147? Else OK. | 4/1/12 (PH) | ? | PROCESSING | ||||
3/23/12(TKN) | 3/23/12(TKN) | 3/26/12 (AG,VF,JN) | 03/29/12(FM) | 03/29/12(FM) | 03/29/12(OM)test 1 | APV0 and APV2 display a large spread in the noise, else OK | 3/31/12(OM) | ? | READY | ||||
3/23/12(TKN) | 3/23/12(TKN) | 3/26/12 (AG,VF,JN) | 03/29/12(FM) | 03/29/12(FM) | 03/29/12(OM) test 1 | OK | 3/31/12 (OM)? | ? | PROCESSING | ||||
3/23/12(TKN) | 3/23/12(TKN) | 3/26/12 (AG,VF,JN) | 03/29/12(FM) | 03/29/12(FM) | 3/30/12(OM) test 1 | Channel 254 bad (as on hybrid), else OK: test 7 shows tails on ch. 48 | 3/31/12(OM) | ? | PROCESSING | ||||
3/26/12(TKN) | 3/26/12(TKN) | 03/27/12(PH) | 03/30/12(FM) | 03/30/12(FM) | Channel 220 unbonded in test 1 and shows tails in test 2: test 2 | 4/1/12 (PH) | ? | PROCESSING | |||||
3/26/12(TKN) | 3/26/12(TKN) | 03/27/12(PH) | 03/30/12(FM) | 03/30/12(FM) | OK: test 2 | 4/1/12(PH) | ? | PROCESSING | |||||
3/26/12(TKN) | 3/26/12(TKN) | 03/27/12(PH) | 03/30/12(FM) | 03/30/12(FM) | Chan 135 and entire APV0 have long baseline distribution tails. Everything else is OK: test 3 | 4/1/12 (PH) | ? | PROCESSING | |||||
3/27/12(TKN) | 3/27/12(TKN) | 03/27/12(PH) | 03/30/12(FM) | 03/30/12(FM) | For 1st 3 tests, had 4 bad channels: | 04/03/12(OM) | ? | PROCESSING | |||||
3/27/12(TKN) | 3/27/12(TKN) | 03/30/12 (AG,VF) | 04/01/12(FM) | 04/01/12(FM) | 04/02/12(OM) | Channel 180 has tails on test 1 only, else OK: test 8 | 04/03/12(OM) | ? | PROCESSING | ||||
3/27/12(TKN) | 3/27/12(TKN) | 03/30/12 (AG,VF) | 04/01/12(FM) | 04/01/12(FM) | 04/02/12(OM) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 | OK; Persistent high noise?? | ? | ? | PROCESSING | ||||
3/30/12 (TKN) | 3/30/12 (TKN) | 4/1/12 (PH) | 04/02/12(FM) | 04/02/12(FM) | 04/03/12(OM) | Tails on channel 170, else OK | 04/03/12(OM) | ? | PROCESSING | ||||
3/30/12 (TKN) | 3/30/12 (TKN) | 4/1/12 (PH) | 04/02/12(FM) | 04/02/12(FM) | 04/02/12(OM) test 1 | OK | 04/03/12(OM) | ? | PROCESSING | ||||
3/30/12 (TKN) | 3/30/12 (TKN) | 4/1/12 (PH) | 04/03/12(FM) | 04/03/12(FM) | A calibration line on APV4 has low offsets. Everything else is OK: test 2 | 04/05/12(OM) | ? | PROCESSING | |||||
4/1/12(TKN) | 4/1/12(TKN) | 4/2/12 (TKN,FM) | 04/03/12(FM) | 04/03/12(FM) | Channel 622 is a pinhole (as on sensor? Sensor data says 623). Channels 203, 300-302 have long pedestal tails (as on hybrid). Else OK. | 04/05/12(OM) | ? | PROCESSING | |||||
4/1/12(TKN) | 4/1/12(TKN) | 4/2/12 (TKN,FM) | 04/03/12(FM) | 04/03/12(FM) | OK | 04/05/12(OM) | ? | PROCESSING | |||||
4/1/12(TKN) | 4/1/12(TKN) | 4/2/12 (TKN,FM) | 04/03/12(FM) | 04/03/12(FM) | Channel 608 has long tails; possibly bad. Else OK. | 04/05/12(OM) | ? | PROCESSING |
To add a half-module, copy the line below into the table and fill in the ?s.
|[Half-module ?] | [Hybrid ?] | [Sensor ?] | ? | ? | ? | ? | ? | ? [test 1|Half-module ?, test 1]| ? | ? | ? |[Module ?]| PROCESSING |