Nominal Operating Voltages and Currents
Experiment=detdaq18, Run=222 (Date)
Pedestal Summary
Method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxx
Gain Mode | Mean Ped Value (ADU) | Number of NCP |
---|
FH | 3268 ± 231 | 4829 |
FM | 3205 ± 225 | 4777 |
FL | 3175 ± 221 | 4758 |
AHL | 3271 ± 229 | 4842 |
AML | 3207 ± 224 | 4797 |
Noise Summary
Gain Mode | Variation (sigma) | Number of NCP, Map |
---|
FH | 6 | 419276 |
FM | 3 | 393224 |
FL | 2 | 402881 |
AHL | 7 | 133023 |
AML | 3 | 383411 |
Gain Mode | Number of Gain NCP, Map |
---|
FH | 1 |
FM | 1 |
FL | 6017 |
AHL | 157 |
AML | 71 |
Linearity (signal injection)
Thermal measurements (all thermometers)
Experiment History
Experiment | Hutch | Shifts | Date Deployed |
---|
Commissioning | XCS | 1 | 15 July 2020 |