Identification Details and Hardware Changelog
Nominal Operating Voltages and Currents
Ch. | Line | V | I |
---|---|---|---|
Performance Measures
Experiment=detdaq18, Run=222 (Date)
Pedestal Summary
Method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxx.
Map of pedestal non-conforming pixels for all gain modes here.
Gain Mode | Mean Pedestal Value (ADU) | Number of NCP | Pixels Map, Histogram |
---|---|---|---|
FH | 3268 ± 231 | 4829 | Graphs |
FM | 3205 ± 225 | 4777 | Graphs |
FL | 3175 ± 221 | 4758 | Graphs |
AHL | 3271 ± 229 | 4842 | Graphs |
AML | 3207 ± 224 | 4797 | Graphs |
Noise Summary
Map of noise non-conforming pixels for all gain modes here.
Gain Mode | Variation (sigma) | Number of NCP | Pixels Map, Histogram |
---|---|---|---|
FH | 6 | 419276 | Graphs |
FM | 3 | 393224 | Graphs |
FL | 2 | 402881 | Graphs |
AHL | 7 | 133023 | Graphs |
AML | 3 | 383411 | Graphs |
Gain and uniformity (photon source)
Map of gain non-conforming pixels for all gain modes here.
Gain Mode | Number of NCP |
---|---|
FH | 1 |
FM | 1 |
FL | 6017 |
AHL | 157 |
AML | 71 |
Linearity (signal injection)
Thermal measurements (all thermometers)
Experiment History
Experiment | Hutch | Shifts | Date Deployed |
---|---|---|---|
Commissioning | XCS | 1 | 15 July 2020 |