Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Identification Details and Hardware Changelog

 


Nominal Operating Voltages and Currents

Ch.LineVI
    





Performance Measures

Experiment=detdaq18, Run=222 (Date)

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as

...

Summary Values

DateMeanStdDev#Bad
    

Pedestal Map

...

xxxx

Gain ModeMean Ped Value (ADU)

Number of NCP

FH3268 ± 2314829
FM3205 ± 2254777
FL3175 ± 2214758
AHL3271 ± 2294842
AML3207 ± 2244797


Noise Summary


Gain ModeVariation (sigma)Number of NCP, Map
FH6419276
FM3393224
FL2402881
AHL7133023
AML3383411


Gain and uniformity (photon source)

...

Gain ModeNumber of Gain NCP, Map
FH

 1

FM 1
FL 6017
AHL157
AML 71

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History

...


ExperimentHutchShiftsDate Deployed
CommissioningXCS115 July 2020