Content
Metrology data
Detector epix10ka2m.0 (old), previous
Kenney, Christopher J.<kenney@slac.stanford.edu>
Dubrovin, Mikhail
Nakahara, Kazutaka;Nayak, Bhavna;Hansson, Conny;
Blaj, Gabriel
HI Mikhail,
Here is the metrology file for repaired ePix10kA 2M
Let me know if you have any trouble with it please.
THanks,
Chris
Measurement map
WARNING: Q0 is in the lower right corner - use parameter -r2
See also: Optical metrology of epix10ka2m.0 from 2020-06-25
Quality check
optical_metrology_check 2024-08-09-epix10ks2m.0-metrology.txt
X-Y quality check for optical metrology measurements
----------------------------------------------------------------------------------------------------------------------------
segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL)
----------------------------------------------------------------------------------------------------------------------------
segm: 0 36704 36758 -750 -696 39700 39704 640 644 1.04328 54078 54116 -37 -54 -4
segm: 1 36687 36677 -756 -766 39682 39725 630 673 1.09806 54030 54102 -72 10 -43
segm: 2 36699 36726 -353 -326 39709 39706 320 317 0.48987 54089 54077 12 -27 3
segm: 3 36705 36714 -661 -652 39727 39699 630 602 0.94708 54101 54082 19 -9 28
segm: 4 36708 36703 613 608 39716 39691 -580 -605 0.88094 54064 54116 -52 5 25
segm: 5 36712 36687 597 572 39725 39692 -559 -592 0.84332 54054 54103 -48 25 33
segm: 6 36708 36724 931 947 39713 39669 -840 -884 1.35524 54091 54079 11 -16 44
segm: 7 36683 36655 626 598 39685 39700 -566 -551 0.88335 54053 54039 13 28 -15
segm: 8 36698 36708 -663 -653 39696 39707 564 575 0.94952 54048 54103 -54 -10 -11
segm: 9 36705 36700 -595 -600 39727 39723 526 522 0.86172 54075 54111 -36 5 4
segm:10 36717 36692 -515 -540 39716 39709 494 487 0.76101 54088 54080 7 25 7
segm:11 36689 36671 -882 -900 39702 39685 851 834 1.28590 54079 54046 33 18 17
segm:12 36715 36714 724 723 39719 39693 -661 -687 1.04390 54091 54092 0 1 26
segm:13 36719 36714 643 638 39660 39709 -639 -590 0.92466 54061 54087 -26 5 -49
segm:14 36674 36675 775 776 39699 39688 -720 -731 1.11926 54051 54060 -9 -1 11
segm:15 36676 36708 699 731 39677 39684 -670 -663 1.03230 54053 54058 -4 -32 -7
----------------------------------------------------------------------------------------------------------------------------
WARNING segm 1: |-72| > 60.0
Z quality check for optical metrology measurements
-----------------------------------------------------------------------------------------------------------------------------------------
segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um)
-----------------------------------------------------------------------------------------------------------------------------------------
segm: 0 36731 39702 36731 39702 412 397 143 128 404 135 15 15 404 135 0.63094 0.19555 15.617
segm: 1 36682 39703 36682 39703 588 376 290 78 482 184 212 212 482 184 0.75282 0.26553 212.135
segm: 2 36712 39707 36712 39707 461 460 310 309 460 309 1 1 460 309 0.71865 0.44658 1.313
segm: 3 36709 39713 36709 39713 550 637 227 314 593 270 -87 -87 593 270 0.92625 0.39026 -87.022
segm: 4 36705 39703 39703 36705 -851 304 569 1724 -273 1146 -1155 -1155 1146 -273 1.65404 -0.42691 1154.824
segm: 5 36699 39708 39708 36699 152 212 445 505 182 475 -60 -60 475 182 0.68535 0.28414 60.471
segm: 6 36716 39691 39691 36716 114 138 406 430 126 418 -24 -24 418 126 0.60338 0.19662 24.392
segm: 7 36669 39692 39692 36669 292 164 519 391 228 455 128 128 455 228 0.65676 0.35625 -127.951
segm: 8 36703 39701 36703 39701 240 476 51 287 358 169 -236 -236 358 169 0.55884 0.24389 -235.914
segm: 9 36702 39725 36702 39725 412 520 233 341 466 287 -108 -108 466 287 0.72743 0.41394 -108.071
segm:10 36704 39712 36704 39712 586 488 189 91 537 140 98 98 537 140 0.83820 0.20199 97.554
segm:11 36680 39693 36680 39693 483 561 226 304 522 265 -78 -78 522 265 0.81533 0.38251 -78.328
segm:12 36714 39706 39706 36714 203 450 446 693 326 569 -247 -247 569 326 0.82173 0.50951 247.276
segm:13 36716 39684 39684 36716 426 331 458 363 378 410 95 95 410 378 0.59265 0.59063 -95.485
segm:14 36674 39693 39693 36674 156 214 519 577 185 548 -58 -58 548 185 0.79096 0.28902 58.133
segm:15 36692 39680 39680 36692 265 113 511 359 189 435 152 152 435 189 0.62808 0.29513 -152.311
-----------------------------------------------------------------------------------------------------------------------------------------
WARNING segm 1: |212.1| > 100.0
WARNING segm 4: |1154.8| > 100.0
WARNING segm 7: |-128.0| > 100.0
WARNING segm 8: |-235.9| > 100.0
WARNING segm 9: |-108.1| > 100.0
WARNING segm 12: |247.3| > 100.0
WARNING segm 15: |-152.3| > 100.0
Fixing z deviation of segment 4
WARNING segm 4: |1154.8| > 100.0
shows that a single point measurement on segmeent 4 deviates >1mm (panel is not flat)
point 2 in this panel has largest deviation of z from others - apply 1.234 mm correction to this point
Quad 1
point X Y Z
1 537063 1045648 390930
#2 536483 1082356 390079 # correction of Z 0079 -> 1234
2 536483 1082356 391234
3 576174 1082964 391803
4 576779 1046261 391499
optical_metrology_check 2024-08-09-epix10ks2m.0-metrology-corr.txt
Z quality check for optical metrology measurements
-----------------------------------------------------------------------------------------------------------------------------------------
segm: SA LA XSize YSize dZS1 dZS2 dZL1 dZL2 dZSA dZLA ddZS ddZL dZX dZY angXZ(deg) angYZ(deg) dz3(um)
-----------------------------------------------------------------------------------------------------------------------------------------
segm: 0 36731 39702 36731 39702 412 397 143 128 404 135 15 15 404 135 0.63094 0.19555 15.617
segm: 1 36682 39703 36682 39703 588 376 290 78 482 184 212 212 482 184 0.75282 0.26553 212.135
segm: 2 36712 39707 36712 39707 461 460 310 309 460 309 1 1 460 309 0.71865 0.44658 1.313
segm: 3 36709 39713 36709 39713 550 637 227 314 593 270 -87 -87 593 270 0.92625 0.39026 -87.022
segm: 4 36705 39703 39703 36705 304 304 569 569 304 569 0 0 569 304 0.82106 0.47452 0.397
segm: 5 36699 39708 39708 36699 152 212 445 505 182 475 -60 -60 475 182 0.68535 0.28414 60.471
segm: 6 36716 39691 39691 36716 114 138 406 430 126 418 -24 -24 418 126 0.60338 0.19662 24.392
segm: 7 36669 39692 39692 36669 292 164 519 391 228 455 128 128 455 228 0.65676 0.35625 -127.951
segm: 8 36703 39701 36703 39701 240 476 51 287 358 169 -236 -236 358 169 0.55884 0.24389 -235.914
segm: 9 36702 39725 36702 39725 412 520 233 341 466 287 -108 -108 466 287 0.72743 0.41394 -108.071
segm:10 36704 39712 36704 39712 586 488 189 91 537 140 98 98 537 140 0.83820 0.20199 97.554
segm:11 36680 39693 36680 39693 483 561 226 304 522 265 -78 -78 522 265 0.81533 0.38251 -78.328
segm:12 36714 39706 39706 36714 203 450 446 693 326 569 -247 -247 569 326 0.82173 0.50951 247.276
segm:13 36716 39684 39684 36716 426 331 458 363 378 410 95 95 410 378 0.59265 0.59063 -95.485
segm:14 36674 39693 39693 36674 156 214 519 577 185 548 -58 -58 548 185 0.79096 0.28902 58.133
segm:15 36692 39680 39680 36692 265 113 511 359 189 435 152 152 435 189 0.62808 0.29513 -152.311
-----------------------------------------------------------------------------------------------------------------------------------------
WARNING segm 1: |212.1| > 100.0
WARNING segm 7: |-128.0| > 100.0
WARNING segm 8: |-235.9| > 100.0
WARNING segm 9: |-108.1| > 100.0
WARNING segm 12: |247.3| > 100.0
WARNING segm 15: |-152.3| > 100.0
Geometry file
optical_metrology_epix10ka2m -i 2024-08-09-epix10ks2m.0-metrology-corr.txt -o 2024-08-09-epix10ks2m.0-geometry.txt -v1 -r2
2024-08-09-epix10ks2m.0-geometry.txt
header
...
# HDR PARENT IND OBJECT IND X0[um] Y0[um] Z0[um] ROT-Z ROT-Y ROT-X TILT-Z TILT-Y TILT-X
CAMERA 0 EPIX10KA:V1 0 59844 -22241 -76 90 0 0 0.48987 0.04112 -0.03980
CAMERA 0 EPIX10KA:V1 1 60431 -62816 -39 90 0 0 0.94708 -0.01010 -0.24504
CAMERA 0 EPIX10KA:V1 2 17933 -22959 44 90 0 0 1.09806 -0.13421 -0.07185
CAMERA 0 EPIX10KA:V1 3 18727 -63726 -56 90 0 0 1.04328 -0.20348 0.05148
CAMERA 0 EPIX10KA:V1 4 -22260 -59944 -27 0 0 0 0.88335 -0.02454 -0.04531
CAMERA 0 EPIX10KA:V1 5 -62928 -60698 14 0 0 0 0.88094 0.13926 0.07415
CAMERA 0 EPIX10KA:V1 6 -23036 -17896 -90 0 0 0 1.35524 -0.08156 -0.19896
CAMERA 0 EPIX10KA:V1 7 -63702 -18688 120 0 0 0 0.84332 0.00433 -0.11709
CAMERA 0 EPIX10KA:V1 8 -60020 22316 19 270 0 0 0.94952 0.15730 -0.12332
CAMERA 0 EPIX10KA:V1 9 -60722 62705 7 270 0 0 0.86172 -0.01226 0.04761
CAMERA 0 EPIX10KA:V1 10 -17940 23051 25 270 0 0 1.28590 0.01371 0.13199
CAMERA 0 EPIX10KA:V1 11 -18702 63696 21 270 0 0 0.76101 0.19982 0.15844
CAMERA 0 EPIX10KA:V1 12 22299 59917 18 180 0 0 0.92466 0.08879 -0.19116
CAMERA 0 EPIX10KA:V1 13 63030 60666 52 180 0 0 1.11926 -0.10826 0.10936
CAMERA 0 EPIX10KA:V1 14 23249 17974 -58 180 0 0 1.04390 -0.13925 -0.11002
CAMERA 0 EPIX10KA:V1 15 63797 18643 24 180 0 0 1.03230 0.05487 0.10462
IP 0 CAMERA 0 0 0 10000 90 0 0 0.00000 0.00000 0.00000
[I] L1023: geometry constants saved in file 2024-08-09-epix10ks2m.0-geometry.txt
ana-4.0.62-py3 [dubrovin@sdflogin001:~/LCLS/ref-detector/alignment/epix10ka2m/calib-xxx-epix10ka2m.0-2024-08-09]$
References