p19-0 ring tubes up interactive map 

Welcome to the interactive map of the first thermomechanical ring prototype at SLAC. You can visit the Cooling Tubes Down map here.

To explore the ring map in interactive mode, use the buttons within the Gliffy diagram macro in the table on the left. You may need to enter full screen. Hover your mouse over the image to get the full screen menu option. Please note that the CMM measurements seem to take an especially long time to display. 

p19-0 ring prototype interactive diagram

Screenshots from the Loading layers of the Interactive Map

Observations, properties, and measurements by Heater ID

Properties

Quads

Heater ID: Quads, batch #1#2#3#4#5#6#7#8#9#10UncertaintySource
Length (cm)4.1104.1064.1024.1014.1054.1014.1014.1084.1054.103±30 μmHeaters.xlsx
Width (cm)4.2544.2454.2444.2464.2534.2544.2644.2544.2554.247±30 μmHeaters.xlsx
Thickness (μm)310310315312312315312312312312±5 μm Heaters.xlsx
Resistance when wires first bonded (Ω)2.4

2.4

2.42.62.42.52.52.32.42.4±0.1 ΩHeaters.xlsx
Resistance: mounted on ring, before spray painting  (Ω)2-3222222222
Heaters.xlsx
Mass (g)







1.4981.564± 0.001 g 
Module thickness at center, measured with micrometers (μm)







305303± 1 μm
Wire thicknesses, measured with calipers (mm)











Triplets

Heater ID: Singles, batch #1#2#3#4#5#6#7#18#9UncertaintySource
Length (cm)








±30 μmHeaters.xlsx
Width (cm)








±30 μmHeaters.xlsx
Thickness (μm)








±5 μm Heaters.xlsx
Resistance when wires first bonded (Ω)2.4

2.4








±0.1 ΩHeaters.xlsx
Resistance: mounted on ring, before spray painting  (Ω)2-322222222
Heaters.xlsx
Mass (g)








± 0.001 g 
Module thickness at center, measured with micrometers (μm)








± 1 μm
Wire thicknesses, measured with calipers ()










Heater visual inspection

Quads

Heater ID#1#2#3#4#5#6#7#8#9#10
Before IPA wash & dry rag polish 









After IPA wash & dry rag polish









Comments







Particles across its surface.

Nico tried to blow them off with an ion gun.

No effect 

Noticed a peculiar speck artifact.

Did not come off with a single IPA cleaning.

Appears to be Cu

Triplets

Heater ID#1#2#3#4#5#6#7#18#9
Before IPA wash & dry rag polish 








After IPA wash & dry rag polish








Comments








Loading experience

Quads

Heater ID#1#2#3#4#5#6#7#8#9#10
Load date







2021-05-032021-05-03
Load position







Q9Q10

Shim testing results

Expected glue layer thickness (μm)









~200100-200

100-μm shim 









PassPass

200-μm shim 









FailPass

Glue batch

Syringe ID

2021-A-142021-A-142021-A-142021-A-142021-A-152021-A-152021-A-152021-A-152021-A-122021-A-12

Dispensing

Estimated glue mass per star (mg)









7171

Averaged mass tests









2, 3, 4, 52, 3, 4, 5 

Time elapsed since syringe fully thawed 









6464

Minutes module held under vacuum (Hold time)









OvernightOvernight

Triplets

Heater ID#1#2#3#4#5#6#7#18#9
Load date








Load position








Shim testing results

Expected glue layer thickness (μm)










100-μm shim 










200-μm shim 










Glue batch

Syringe ID










Date dispensed










Dispensing

Estimated glue mass per star (mg)










Averaged mass tests










Time elapsed since syringe fully thawed 










Minutes module held under vacuum (Hold time)










Metrology

Here is the link to the z measurements of the tube-up side of the 19-0 ring: https://docs.google.com/spreadsheets/d/1-wvzPj9BKTd7p5U0iAwyJC_zR2qVB3EyFneJkcj2nHE/edit?usp=sharing

Thermal imaging