Power integration - ePixHrM electric cable specs

The cable goes from the detector box to the power supply and external monitoring instruments. It must feature a DBM25P032N on one side and BNC / bananas on the other side.

DBM25S032N Pins OUT - Male to Female

DBM25P032N

Signal name

HD 25 pins female

1

MPS

BNC

2

DGND

Banana

3

TG

BNC

4

DGND


5

DAQ

BNC

6

DGND


7

RUN

BNC

8

DGND


9

TTL_OUT

BNC

10

DGND


11

TTL_IN

BNC

12

DGND


13

THERM_IN

Banana

14

DGND


15

THERM_OUT

Banana

16

DGND


17

AGND

Banana

18

AGND


19

ANA_24V

Banana

20

ANA_24V


21

DIG_24V

Banana

22

DIG_24V


23

DGND


24

HV_GND

Banana

25

HV_BIAS

Banana




On the DBM25P032N side, the following pins must be connected together:

·         Pins 2,4,6,8,10,12,14,16,23

·         Pins 17 and 18

·         Pins 19 and 20

·         Pins 21 and 22


Cable length: 5.5ft


Twit cables used for:

·         DGND and DIG_24V

·         AGND and ANA_24V

·         HV_GND and HV_BIAS


DBM25S032N Connector


Final product

                      DBM25P032N  side                                                    Banana & BNC side


Data and timing cables


Necessary hardware for connecting data and timing

  • One 24 channel type B fiber MTP/MPO optics cable . If the detector is in vacuum, a fiber feed through needs to be used
  • Multimode to Single mode conversion box
  • 4 40Gbps SM or MM transceivers for data. Either one or another,
  • 3 MTP/MPO Fiber optics cables
  • 1 40Gbps SM or MM transceiver for timing
  • MTP fiber breakout



The detector will be connected to the MM to SM converter box (input LEAP0) using the 24 channel type A/B cable. LEAP0 LANE[3:0] and LANE[7:4] ports will be connected to the data acquisition server using two FO cables and 4 transceivers. LANE[11:9] (namely lane 11) will be connected to the LCLSII timing system using the MTP fiber breakout, remaining transceiver and cable. The MM to SM converter box has to be as close as possible to the detector.


If a type B cable is used, you need a straight coupler (both keys of both fiber optics cables are on one side). 

If a type A cable is used, you need a cross coupler (one key on one side, and the other on the other side).


Detector attachment to the cooling system

To connect the detector to the cooling system, use the hardware mentioned here.