This page is intended to serve as a cheat sheet documenting trigger settings for common devices.
N.B. Cspad numbers were determined in CXI and worked in XPP.
Febo numbers were determined in XCS.
Camera Model |
Pulse Polarity |
Pulse Delay |
Pulse Width |
comments |
---|---|---|---|---|
Adimec Opal 1000 |
Pos |
0.0008 |
0.0002 |
Pulse width affects exposure time. 0.0001 sec is too short. |
Adimec Opal 4000 |
Pos |
0.0008 |
0.0002 |
Maximum trigger rate is 30Hz. |
Pulnix TM-6740CL |
Neg |
0.0006 |
0.0005 |
Pulse width affects exposure time. |
LBL FCCD |
Pos |
|
|
Pulse width does not affect exposure time. 60 Hz strongly recommended. |
Cspad |
Neg |
5.59092e-4 |
1e-5 |
Ryan says will change to Pos polarity during shutdown |
Cspad2x2 |
Pos |
5.58487e-4 |
1e-5 |
Critical that the "Run" trigger coming from controls evr have polarity |
Febo (normal) |
Pos |
7.78e-4 |
1e-6 |
For normal mode, trigger width not critical |
Febo (Average) |
Pos |
4.41e-4 |
1e-6 |
For averaging mode, trigger width not critical |
IPIMB |
Neg |
0.000651824 |
5e-06 |
|
Camera Model |
Pulse Polarity |
Pulse Delay |
Pulse Width |
comments |
---|---|---|---|---|
Timepix |
Neg |
0.00085 |
0.0002 |
Pulse width affects exposure time. |
Model |
Pulse Polarity |
Pulse Delay |
Pulse Width |
comments |
---|---|---|---|---|
General Standards GSC16AI32SSC |
Pos |
|
|
Pulse width is not critical. |