This serves as a guide for moving the Jungfrau detector off and onto the DSA location.
Step-by-step guide
NOTE: It is good practice to check that vacuum levels are at atmospheric conditions (760 torr) prior to disconnecting flanges. This should be checked on the control screens and manual gauges.
- Disconnect both 14" flanges on each end of the Jungfrau. The screws will act as guide posts for pulling back the bellows. Note: There were interference issues with the Optique Peter, on the top 4 screws of the flange, located in the +z direction.
- Slowly pull back the bellows and secure the 3 handles onto the lead screws and foil off all vacuum side components. This requires a minimum effort of 2 people.
Crane preparation for the 14" flange+stand:
- Remove all 4 bolts securing the stand to the platform. Note: Consult the CXI crane book for the most current lift plan and check for completion of the daily crane inspection.
Removal of the 2 Turbos:
- Remove the flange sitting on top of the 2 turbos to gain access to the bracket holder port. This secures the umbilical cord located inside the Jungfrau.
2. 3 silver plated bolts are removed to disengage the bracket holding the umbilical cord. These are no longer capture and release screws.
This guide is still in progress.