Step-by-step guide - Femtosecond (fs) timing in sample chamber 1 (SC1) at CXI.

Setup

Check all diodes at the beginning of the shift (ideally this should also be done a few days before the experiment to give you time to fix things that don't work), by plugging them into Lecroy and visual check (with bias-T).

Install a fast diode and mirror (roughly at 45 degrees with respect to the beam) on the SC1 post_sample_x stage (CXI:SC1:MMS:02.RBV) downstream on the interaction region. There is a danger of driving this stage into the Jungfrau or other components in SC1, so it might be disabled. It is also good practice to disable the motor when not in use.

Roughly align the fast diode and mirror using the reference laser at the time tool.

Install a regular diode connected to the Acqiris on the sample chamber door and align the reflected reference laser beam on the mirror onto that diode.

Repeat with the fs laser and maximize the Acqiris signal by moving the diode on the chamber door.  An ND filter may be necessary to reduce the signal, as shown in the side-on view below

Also install a fs laser diagnostic diode (OPC-TP) and connect it to the Acqiris. This is useful to confirm that the fs laser is firing during the experiment.

Insert a nozzle rod with both a clear YAG and a frosted YAG.

If doing the timing with the shroud on, the bracket needs to be facing in the -Z direction when inserted into the sample chamber or it will collide with the exit cone.

Determine the position of the X-ray spot at the IP using the clear YAG and the beam with 1x10-4 Transmission (although the beam can be seen with less intensity). Mark the position of the beam with crosshairs (cyan and magenta in the example image below).

Rough timing

If timing is desired to be done at atmospheric pressure in SC1, then the MPS interlock must be bypassed to allow the beam to come to CXI with the detector gate valve close. this requires bypassing Link Node 40, Card 3 Channel 15 to open for the duration of the timing measurement at atmosphere. This bypass can only be enabled by ACR and you must call them to request this at x2151.

Move the fast diode into position (at this point should only need to move the SC1 sample_x stage) and move the IP YAGs out of the beam.

Measure the laser signal on the diode using the LeCroy scope to make sure there is overlap

Shutter the fs laser and unshutter the X-rays. Move the diode around in x and y to maximize the X-ray signal (if the steps are very large then the laser may need to be re-pointed)

Click Store on the Lecroy to save a trace of the X-rays on the scope

Shutter X-rays off and unshutter fs laser.

Temporally overlap the X-rays and the fs laser.

Fine timing

Once the laser is roughly timed, it is time to move onto the finer timing measurement using a cross-correlation measurement.

Background Information

Align the laser with crosshairs on the inline cameras that represent the location of the X-ray beam.

Enlarge the X-ray spot size using DG2 lenses.

Find t0 at the IP with a fine timing scan.