- Detector running and stabilized for two hours, nominal settings.
Record thermometer and humidity sensor readings:
Sensor Airbox T 39.64C Airbox Humidity 0.96% Airbox Pressure 1007 hPa Dewpoint -25.4C Chiller Flow 2.3 l/m Chiller Setpoint 15C
...
Electrical
- Document grounding scheme: Heavy clip ground wire to power supply chassis, nothing else visible:
- Read/record all available voltages and currents
Power Supply | |
---|---|
Temp | 29.3C |
Voltage | 11.792V |
Current | 25.509A |
-
Performance
- Force high gain mode
- Collect 1000 darks for analysis (Conny test 1; pedestals, noise, bad pixel map)
- Force medium gain
- Collect 1k darks (C2, pedestals/offsets, noise, bad pixel map)
- Force low gain
- Collect 1k darks (C3, pedestals/offsets, noise, bad pixel map)
- Create pedestals with 'makepeds -J'
- Offsets due to ambient visible light
...