Identification

Firmware

xxxxupdated date 
Module identifiers   
    

Jungfrau 4M Module Layout

 

Verify Nominal Operating Voltages and Currents

Environmental

Test conditions:

Electrical

Power Supply

 
Temp

29.3C

Voltage

11.792V

Current25.509A

 

Performance

 

Day two was mostly washed out by inability to run the detector due to DAQ config error.  Did verify timing delay or 231us.

 

Day 3:  Looking to generate enough flux per shot to trigger autorange in the JF.  Problem is that fluorescence foils burn through immediately and we see greatly reduced flux.  A script has been written to reposition the foil once per shot so that we're continuously using new foil.

 

 

Plan to go:

 

-        Offsets due to ambient visible light

-        Mechanical

-        Assess

 

 

1)     Beam set up (12h, full shift) (wave8 monitoring throughout)                                      (t = 0)

 

-        No sample (60 min)

-        Assess

 

-        Sample setting (TBD)                                                                                                               (t = 1h)

 

-        Set attenuator for moderate-flux level and scan in z (60 min)                                      (t = 5.5h)

 

-        High gain flats (fixed gain)                                                                                                                     (t = 6.5h)

                                                    i.     Vary bias +/- 20%

 

-         Medium gain flats (fixed gain)                                                                                                             (t = 9h)

                                                    i.     Vary bias +/- 20%

 

-        Low gain flats (fixed gain)                                                                                                                      (t = 11.5h)

                                                    i.     Vary bias +/- 20%

 

-        Full range flats (autoranging)                                                                                                               (t = 14h)

                                                    i.     Vary bias +/- 20%

 

-        Repeat baseline tests      (3h)                                                                                                                 (t = 17h)

 

-        End                                                                                                                                                              (t = 20h)

 

Performance Measures

Pedestal

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxxx.

Summary Values

DateMeanStdDev#Bad
    

Pedestal Map

Noise map

Gain and uniformity (photon source)L

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History

 

ExperimentHutchShiftsDate Deployed
CommissioningMFX115 June 2020

Calibration Records

 

DateHutchDateRecord linkComment
CommissioningMFX7/29/2015 June 2020