Single photon gain/noise for old and new settings (Zongde)

  • ASIC3 gain&noise

Image ModifiedImage Modified

...

  • ASIC2 gain&noise

Image Modified Image Modified

...

Image Added

Cross-talk correction for single photon under mask column-wise (Zongde)


  • Dark pixel common mode correction

Image Added

Image Added

Image Added


...

  • Dark pixel but with low flux photon leakage

Image Added

Image Added


Pedestal and offsets at transition point for new settings for both temperatures (Bhavna/Philip) 



TID lab Charge injection tests (Alex&Xavier)

Charge injection with variable number of columns, while monitoring a control pixel (non-injected), in auto gain. Injected columns are: 14 - 374, 86 - 374, 158 - 374, 230 - 374 and 302 - 374. The cross-talk magnitude increases with the amount of charge injected:

Image Added

...

Same as above but in HDR (high dynamic range) auto gain:
CompTH_ePixM = 24
Precharge_DAC_ePixM = 50
RefinN = 7
RefinP = 7

The apparent step-function point at around 400 pulser value corresponds to gain switching point in injected pixels. Now the question is what causes this jump in non-injected pixels

Image Added

...


Measuring the effect of charge injection block position on the control pixel shows that the step point position (pulser value at which the jump occurs) changes, as well as the baseline after the step. In this measurement the injected columns were 14 - 86, 86 - 158, 158 - 230, 230 - 302 and 302 - 374 (72 columns per injection step), while monitoring the same pixel 103,4.

Image Added

Here is a plot of the first and last extremes for clarity (colors do not correspond to the plot above) 

Image Added

So, not only the amount of charge affects the baseline (non-injected pixel), but the position of the charge on the asic relative to the control pixel.

...

Auto gain (Low noise), Blue trace is injected pixel, orange trace is non-injected pixel I monitor for crosstalk. 

Image Added

Auto gain (High dynamic range)

Image Added


Image Added

Register optimization based on charge injection, part 1 (Alex)

View file
nameAlex_reg_scan1.pdf
height250



Commom mode correction summary:

View file
nameSummary of Investigation on Common mode correction.pptx
height250