Identification Details and Hardware Changelog


Nominal Operating Voltages and Currents

Ch.LineVI




Performance Measures

Experiment=detdaq18, Run=521 (Jan 28, 2020)

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH3215 ± 16523Graphs
FM3157 ± 16535Graphs
FL3157 ± 164 54Graphs
AHL3216 ± 16523Graphs
AML3175 ± 16536Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH614542Graphs
FM23135Graphs
FL21397Graphs
AHL76377Graphs
AML35892Graphs


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH0
FM0
FL480
AHL16
AML9

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History


ExperimentHutchShiftsDate Deployed