Identification Details and Hardware Changelog

FirmwareCarrierID0CarrierID1DigitalCardID0DigitalCardID1AnalogCardID0AnalogCardID1Date
3925999620099657958505536481381232098304122164173926502515213976200215April 2022

0xea020004 = 3925999620


Performance Measures

Summary

For Pedestals, method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.


exp:run (mm/dd/yy)

Temp (oC)Humidity (%)Mean Pedestal Value (ADU) Number of Ped NCPPedestal GraphsNoise (ADU)Number of Noise NCPNoise GraphsConfigRun Notes

detdaq20, run 128 (3/28/22)

16.544118 ± 22632

Histogram/Pixel Map 

Pedestal NCP Map

2539

Histogram/Pixel Map 

Pedestal NCP Map

image
detdaq19, run 263 (11/11/21)63.73922 ± 22329

Histogram/Pixel Map 

Pedestal NCP Map

2720

Histogram/Pixel Map 

Pedestal NCP Map

image
detdaq18, run 267 (9/3/20)

3908 ± 22426

Histogram/Pixel Map 

Pedestal NCP Map

193332

Histogram/Pixel Map 

Pedestal NCP Map

image

Thermal measurements (all thermometers)

Configuration Settings

History

Run 19

Experiment

Hutch

Shifts

Date Deployed (mm/dd/yy)

LW64XPP
9/22/21


Testing 

Where

Who

Why

When (mm/dd/yy)

Notes
TIDDoeringTest LCLS-II firmware12/06/21All clear