Attendance:


Presentation:


Agenda/Notes:

  • Any testing or data analysis update
    • ePix100 Gain calibration (finished data collection 2 out of 7)
      • setup with Cu target
      • Single photon spectrum
        •   
      • Philip has made the Sbatch script working, so we can assign the data analysis to  compute nodes.
    • ePixM units status–Xavier
      • In total: 8 new ePixM units with 4 ASICs, 3 old detectors (1 out of 3 died during the beamtime), so 2 old working units now.
      • 8 New units with 4 ASICs: 
        • RX003_4ASICs: 3 ASICs work
        • RX004_4ASICs: 4 ASICs work but quite a few lanes don't work on one ASIC (best detector so far)
        • RX006_4ASICs: 3 ASICs work, one ASIC gives strange pictures with the laser  

        • RX008_4ASICs: 3 ASICs work, one doesn't

        • RX009_4ASICs → problem when powering the board! Cannot test it. 


        →I still need to test the 3 remaining ones. 


    • ASC setup development – Zongde
      • Barium (K-alpha: 32.2kev) target ready to use.
      • Terbrium target just received, will do cleaning procedure tomorrow.
  • No labels