Why do I think that 2020-02-25 optical metrology is good

  • very good quality check
  • all z-measurements are around 1 feet, have a lit of repeatable numbers - can't be trusted
Quality check for original optical metrology from 2020-02-25
X-Y quality check for optical metrology measurements 
----------------------------------------------------------------------------------------------------------------------------
segm:        S1      S2     dS1     dS2        L1      L2     dL1     dL2    angle(deg)   D1      D2      dD   d(dS)   d(dL)
----------------------------------------------------------------------------------------------------------------------------
segm: 0   36697   36688      18       9     39685   39676       9       0   -0.01949   54057   54032      24       9       9
segm: 1   36672   36696      90     114     39690   39688     -93     -95   -0.14725   54046   54045       0     -24       2
segm: 2   36703   36700     202     199     39695   39782    -183     -96   -0.28908   54127   54060      67       3     -87
segm: 3   36683   36700     260     277     39677   39695    -243    -225   -0.38763   54060   54039      20     -17     -18
segm: 4   36710   36693    -274    -291     39685   39687     270     272   -0.40785   54063   54049      14      17      -2
segm: 5   36686   36679    -408    -415     39699   39688     406     395   -0.59396   54065   54036      29       7      11
segm: 6   36684   36676    -134    -142     39693   39693     115     115   -0.19920   54036   54055     -18       8       0
segm: 7   36696   36705    -287    -278     39693   39706     256     269   -0.40771   54066   54064       1      -9     -13
segm: 8   36675   36684     162     171     39688   39674    -146    -160   -0.24041   54037   54036       1      -9      14
segm: 9   36679   36678     232     231     39671   39683    -220    -208   -0.33429   54034   54034       0       1     -12
segm:10   36679   36682     253     256     39677   39694    -240    -223   -0.36743   54044   54039       5      -3     -17
segm:11   36693   36701     367     375     39714   39690    -343    -367   -0.53539   54057   54075     -17      -8      24
segm:12   36677   36682    -378    -373     39671   39686     340     355   -0.54221   54038   54037       0      -5     -15
segm:13   36670   36681    -108     -97     39685   39679      98      92   -0.14800   54035   54034       0     -11       6
segm:14   36683   36671     -96    -108     39693   39664     104      75   -0.14729   54029   54036      -7      12      29
segm:15   36686   36677    -163    -172     39690   39677     169     156   -0.24184   54046   54034      11       9      13
----------------------------------------------------------------------------------------------------------------------------
  WARNING segm  2:  |67| > 60.0
  WARNING segm  2:  |-87| > 60.0

DAQ panel indexes in optical metrology possible detector orientations

DAQ panel indexes in optical metrology possible orientations
        rot = 0  ^y
        Q0  1  3 |  4  5  Q1
            0  2 |  6  7
        ---------+----------> x
           15 14 | 10  8
        Q3 13 12 | 11  9  Q2

        rot = 1  ^y
        Q3 13 15 |  0  1  Q0
           12 14 |  2  3
        ---------+----------> x
           11 10 |  6  4
        Q2  9  8 |  7  5  Q1

        rot = 2  ^y
        Q2  9 11 | 12 13  Q3
            8 10 | 14 15
        ---------+----------> x
            7  6 |  2  0
        Q1  5  4 |  3  1  Q0

        rot = 3  ^y
        Q1  5  7 |  8  9  Q2
            4  6 | 10 11
        ---------+----------> x
            3  2 | 14 12
        Q0  1  0 | 15 13  Q3

 

Images of panels in data for all detector orientations in metrology processing v1 without quads

  • v1 of the metrology processing script represents simplest approach - each panel coordinates are taken as is to insert in the detector
  • data from mfxc00118 run44 silver-behenate sample

 

Summary: 

  • numeration of quads at rotation looks symmetric - so seems right
  • one orientation of the detector gives correct image and it shows smallest offsets between panels in quads

mfxc00118 run44 silver-behenate

mfxc00318 run 10,13 lysozyme