Why do I think that 2020-02-25 optical metrology is good
- very good quality check
- all z-measurements are around 1 feet, have a lit of repeatable numbers - can't be trusted
Quality check for original optical metrology from 2020-02-25
X-Y quality check for optical metrology measurements ---------------------------------------------------------------------------------------------------------------------------- segm: S1 S2 dS1 dS2 L1 L2 dL1 dL2 angle(deg) D1 D2 dD d(dS) d(dL) ---------------------------------------------------------------------------------------------------------------------------- segm: 0 36697 36688 18 9 39685 39676 9 0 -0.01949 54057 54032 24 9 9 segm: 1 36672 36696 90 114 39690 39688 -93 -95 -0.14725 54046 54045 0 -24 2 segm: 2 36703 36700 202 199 39695 39782 -183 -96 -0.28908 54127 54060 67 3 -87 segm: 3 36683 36700 260 277 39677 39695 -243 -225 -0.38763 54060 54039 20 -17 -18 segm: 4 36710 36693 -274 -291 39685 39687 270 272 -0.40785 54063 54049 14 17 -2 segm: 5 36686 36679 -408 -415 39699 39688 406 395 -0.59396 54065 54036 29 7 11 segm: 6 36684 36676 -134 -142 39693 39693 115 115 -0.19920 54036 54055 -18 8 0 segm: 7 36696 36705 -287 -278 39693 39706 256 269 -0.40771 54066 54064 1 -9 -13 segm: 8 36675 36684 162 171 39688 39674 -146 -160 -0.24041 54037 54036 1 -9 14 segm: 9 36679 36678 232 231 39671 39683 -220 -208 -0.33429 54034 54034 0 1 -12 segm:10 36679 36682 253 256 39677 39694 -240 -223 -0.36743 54044 54039 5 -3 -17 segm:11 36693 36701 367 375 39714 39690 -343 -367 -0.53539 54057 54075 -17 -8 24 segm:12 36677 36682 -378 -373 39671 39686 340 355 -0.54221 54038 54037 0 -5 -15 segm:13 36670 36681 -108 -97 39685 39679 98 92 -0.14800 54035 54034 0 -11 6 segm:14 36683 36671 -96 -108 39693 39664 104 75 -0.14729 54029 54036 -7 12 29 segm:15 36686 36677 -163 -172 39690 39677 169 156 -0.24184 54046 54034 11 9 13 ---------------------------------------------------------------------------------------------------------------------------- WARNING segm 2: |67| > 60.0 WARNING segm 2: |-87| > 60.0
DAQ panel indexes in optical metrology possible detector orientations
DAQ panel indexes in optical metrology possible orientations
rot = 0 ^y Q0 1 3 | 4 5 Q1 0 2 | 6 7 ---------+----------> x 15 14 | 10 8 Q3 13 12 | 11 9 Q2 rot = 1 ^y Q3 13 15 | 0 1 Q0 12 14 | 2 3 ---------+----------> x 11 10 | 6 4 Q2 9 8 | 7 5 Q1 rot = 2 ^y Q2 9 11 | 12 13 Q3 8 10 | 14 15 ---------+----------> x 7 6 | 2 0 Q1 5 4 | 3 1 Q0 rot = 3 ^y Q1 5 7 | 8 9 Q2 4 6 | 10 11 ---------+----------> x 3 2 | 14 12 Q0 1 0 | 15 13 Q3
Images of panels in data for all detector orientations in metrology processing v1 without quads
- v1 of the metrology processing script represents simplest approach - each panel coordinates are taken as is to insert in the detector
- data from mfxc00118 run44 silver-behenate sample
Summary:
- numeration of quads at rotation looks symmetric - so seems right
- one orientation of the detector gives correct image and it shows smallest offsets between panels in quads
mfxc00118 run44 silver-behenate
mfxc00318 run 10,13 lysozyme