Identification Details and Hardware Changelog
Nominal Operating Voltages and Currents
Ch. | Line | V | I |
---|---|---|---|
Performance Measures
Experiment=detdaq18, Run=329 (Date)
Pedestal Summary
Method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxx.
Map of pedestal non-conforming pixels for all gain modes here.
Gain Mode | Mean Pedestal Value (ADU) | Number of NCP | Pixels Map, Histogram |
---|---|---|---|
FH | 2859 ± 140 | 10 | Graph |
FM | 2828 ± 136 | 10 | Graph |
FL | 2816 ± 136 | 10 | Graph |
AHL | 2861 ± 138 | 10 | Graph |
AML | 2831 ± 137 | 10 | Graph |
Noise Summary
Map of noise non-conforming pixels for all gain modes here.
Gain Mode | Variation (sigma) | Number of NCP | Pixels Map, Histogram |
---|---|---|---|
FH | 5 | 425 | Graph |
FM | 2 | 94 | Graph |
FL | 2 | 635 | Graph |
AHL | 6 | 297 | Graph |
AML | 3 | 122 | Graph |
Gain and uniformity (photon source)
Map of gain non-conforming pixels for all gain modes here.
Gain Mode | Number of NCP |
---|---|
FH | 0 |
FM | 0 |
FL | 48 |
AHL | 8 |
AML | 6 |
Linearity (signal injection)
Thermal measurements (all thermometers)
Experiment History
Experiment | Hutch | Shifts | Date Deployed |
---|---|---|---|
Commissioning | XCS | 1 | 15 July 2020 |