This page is intended to serve as a cheat sheet documenting trigger settings for common devices.

CameraLink and PGP Cameras

N.B. Cspad numbers were determined in CXI and worked in XPP.

Febo numbers were determined in XCS.

Camera Model

Pulse Polarity

Pulse Delay

Pulse Width

comments

Adimec Opal 1000

Pos

0.0008

0.0002

Pulse width affects exposure time. 0.0001 sec is too short.

Adimec Opal 4000

Pos

0.0008

0.0002

Maximum trigger rate is 30Hz.

Pulnix TM-6740CL

Neg

0.0006

0.0005

Pulse width affects exposure time.

LBL FCCD

Pos

 

 

Pulse width does not affect exposure time. 60 Hz strongly recommended.

Cspad

Neg

5.59092e-4

1e-5


Cspad2x2
Version 1.5
and Daq release
after 6.1.1

Pos

5e-4

1e-5

Run trigger the same, 5e-4 seconds, or 59524 ticks in the Daq system.
For run triggers coming from controls use 58420 ticks for event code 40,
since the controls triggers time from the event code, not event code 140.
Start with 11020 runTriggerDelay in the configuration
Scan Run Trig Delay in units of 8e-9 seconds to time in or scan across.

N.B. the above assumes a Acqdelay of 280.  AcqDelay is in units of
1.024e-6 seconds and is in series with the two numbers above.

Febo (normal)

Pos

7.78e-4

1e-6

For normal mode, trigger width not critical

Febo (Average)

Pos

4.41e-4

1e-6

For averaging mode, trigger width not critical

IPIMB

Neg

0.000651824

5e-06

 

Other Cameras

Camera Model

Pulse Polarity

Pulse Delay

Pulse Width

comments

Timepix

Neg

0.00085

0.0002

Pulse width affects exposure time.

Other Devices

Model

Pulse Polarity

Pulse Delay

Pulse Width

comments

General Standards GSC16AI32SSC
(Beam Rate ADC)

Pos

 

 

Pulse width is not critical.

  • No labels