Identification Details and Hardware Changelog


Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad (non-conforming pixels aka NCP). 

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram

FH

Graphs
FM

Graphs
FL

Graphs
AHL

Graphs
AML

Graphs


Noise Summary

Gain Mode

Variation (ADU)

Number of NCP

Pixels Map, Histogram

FH

Graphs
FM

Graphs
FL

Graphs
AHL

Graphs
AML

     Graphs     

Bad Pixel Map

        Map all pixels with a non-zero bad status bit, which is defined here.   

Configuration Settings:x


Experiment History:

RunExperimentHutchShiftsDate DeployedNotes 
22L-10323XPP5Dec 15, 2023

X-10103XCS3

 



X-10120XCS2




X-10004XCS5



23L-10351 Mara XCS5

 



L-10023 SchellerXPP5

 



L-10490 FollmerXCS4

 



P-10033 PaleyMFX3

 



L-10389 Guenther MFX4



Notes

9/10/2024 - Tim mentioned the replaced asic is fine, but now noise across camera is not flat, higher than before. Waiting for him to elaborate. 

: Tripper was manually turned off during MFX experiment and we see a large increase in bad pixels in status map. Over 24 hours, some damage has annealed. Also investigating the tilted geometry in MFX.  


  • No labels