Attendance:
Presentation:
Agenda/Notes:
- Any testing or data analysis update
- ePix100 Gain calibration (finished data collection 2 out of 7)
- Philip has made the Sbatch script working, so we can assign the data analysis to compute nodes.
- Data collection (7 epix100 units done) , etch 30 unit may need redo.
- ePix100 Gain calibration (finished data collection 2 out of 7)
-
- ePixM units status–Xavier
- In total: 8 new ePixM units with 4 ASICs, 3 old detectors (1 out of 3 died during the beamtime), so 2 old working units now.
- 8 New units with 4 ASICs:
- RX003_4ASICs: 3 ASICs work
- RX004_4ASICs: 4 ASICs work but quite a few lanes don't work on one ASIC (best detector so far)
RX006_4ASICs: 3 ASICs work, one ASIC gives strange pictures with the laser
RX008_4ASICs: 3 ASICs work, one doesn't
RX009_4ASICs → problem when powering the board! Cannot test it.
→I still need to test the 3 remaining ones. - HV biasing Voltage determination (Presentation)
- Recommended voltage ~-80V. (-60-70 should also be fine).
- Burn test over the weekend at -85V. Check to see that current draw is stable.
- ePixM units status–Xavier
- ASC setup development – Zongde