Procedure to start up and align the IR laser in the laser room.
Assumption is that the oscillator and Regen are running.
Preparations
- Check status of chillers and refill if necessary:
- Outside: Regen
- Inside: MPA crystal, Oscillator etc.
- Chiller load for SAGAs: https://docs.google.com/document/d/1tVj5wvT-RQI_wx3ykWPv121MkaEuvCQ05LFtRqwuRRc/
- Check for any other alarms. Regen and oscillator running.
- Check beam out of MPA is blocked with power meter
Start up/warm up
- Turn on PreAmp (Flashlamp and Q-Switch)
- Turn on both SAGAs (shutters closed)
Alignment
- Open auto align GUI (facet home → Matlab GUIs → S20 Laser Tools → S20 Laser Auto Alignment)
- Open S20 Laser Room Multi Profile Monitor (facet home → Matlab GUIs → S20 Laser Tools → S20 Laser Room Multi Profile Monitor)