YFE Turn on Procedure:

  • Set hutch to “Class 1”, “Class 4 Glass Laser” or “Class 4 Dual Laser”or “Class 4 Split Mode
  • Don the appropriate eyewear
  • At the LSS display, badge in and open Shutter S4


Turn on each of the 4 chillers:

  • Requires a ~ 30 minute run time to reach stable set temp
  • YFE output energy will be low until set temperature is reached
  • Set Temperature for each chiller is:
  • 2mm ABCD = 30.3 degrees C with 49psi and a .95GPM flow rate
  • 2mm EFGH = 30.3 degrees C with 49psi and a .95GPM flow rate
  • 6mm AB = 26 degrees C with 43psi and a 2.0GPM flow rate
  • 10mm CD = 24 degrees C with 47psi and a 2.0GPM flow rate