Contents:


Process and current progress:


Requirements to meet:


RequirementsMeasured resultnote
Mode of operationAHL (functional)Verified

AML (functional)Verified

FH (functional)Verified

FM (functional)Verified

FL (functional)Verified
Pixel size100x100 um2Verified
Range

Autoranging: >40000 keV

(transitions 400 keV/1200 keV selectable)




FH >400 keV


FM >1,200 keV


FL > 40,000 keV

Noise r.m.s.

<~400eV (110 e- rms)



Note that the specifications must meet the requirements at a readout rate that exceeds 5102 Hz or 4644 Hz.


Calibration Routine:

Calibration constants and how we determine them (TBD):

Pedestals:

Gain modeMeasurementCalculationNote
FHDirect measurement of darkaverage of 1000 events?Takepeds
FMDirect measurement of darkaverage of 1000 events?Takepeds
FLDirect measurement of darkaverage of 1000 events?Takepeds
AHL-HDirect measurement of darkaverage of 1000 events?Takepeds
AHL-L


AML-MDirect measurement of darkaverage of 1000 events?Takepeds
AML-L



Gains:

Gain modeMeasurementCalculationNote
FHSingle photon data
Single photon data?
FMFH single photon gain * linearity gain ration(FH/FM)?

FLFH single photon gain * linearity gain ration(FH/FL)? 

AHL-HFH Gain?
is FH and AHL_H gains the same
AHL-LFH single photon gain * linearity gain ration(FH/FL)? 
is FH and AHL_H gains the same
AML-MFH single photon gain * linearity gain ration(FH/FM)? 
is FH and AHL_H gains the same
AML-LFH single photon gain * linearity gain ration(FH/FL)? 
is FH and AHL_H gains the same


Proposed calibration scheme:


Results (follow the link down the rabbit hole):

ePixHR results page





  • No labels