Goal of this procedure: Enter oven mode and reach desired oven profile

Instructions: To keep track of changes to the procedure - copy this page, date it, and add execution notes in red.

Table of Contents:


Procedure

Checkout and setup


Procedure Execution notes

Record goal density and helium buffer pressure


Set 10 Torr gauge set points to plus and minus 1 Torr of desired pressure

  • Note: the hysteresis value is where the trip occurs, the lower value is where the fault will clear
 

Set the 1000 Torr gauge setpoint to 20 Torr


Confirm helium gas bottle is connected to fill line 1, regulator set to 5-10 psig 



Procedure for oven turn on