For Naming Czars:

We have two decarad systems. Each consists of 10 small Geiger-Mueller (GM) radiation detecting heads that are connected (signal & high voltage (HV)) to a central unit which has some brains to control the HV level and read out the measured dose rates. The plan is to mount these to two frames that will be positioned near cryomodules during RF commissioning to measure radiation. The systems will be moved every day or two as we commission all cavities. The systems were designed and built at JLab and used for many years. The decarads are talked to through a DB9 connector connected to a Digi Portserver. The software uses EPICS ASYN and Stream Device to send the commands to set the HV or read back the measurements. The software is part of the RadiationMonitoring slac epics package. The proposed software is currently in afs:

~jnelson/sandbox/RadiationMonitoring/RadiationMonitoring-git/RadiationMonitoringApp/Db/common:
decaradChannel.template - records associated with each GM head
decaradCommon.template - records associated with the system
decarad.substitutions - substitution file for templates

In ~jnelson/sandbox/RadiationMonitoring/RadiationMonitoring-git/iocBoot/sioc-sys0-rm01 is the st.cmd file.

Below is a description of the PVs with proposed names

Proposed PV names:

PV NamePurposeSave/restoreArchive
$(P):FIRMread firmware number

$(P):HRTBTread heartbeat
60s
$(P):HVSTATread status of high voltage
60s
$(P):TEMPread temperature of chassis
60s
$(P):HVMONhigh voltage reading (V)
60s
$(P):HVSETPThigh voltage setting (V)

yes

60s
$(P):RACK:PositionString to hold a description of the rack positionyes
$(P):nn:DoseRateread back of dose from GM tube nn (01-10) in mR/hr (close enough, it saturates at xxx)
60s
$(P):nn:Positionstring to hold a description of the position of GM tube number nnyes
$(P):SEQsequence to readout the PVs at 1 Hz. The SEQ record has a delay of 0.06 seconds between sending 1 to each PV's PROC field so commands don't collide in the cable

There are two sets of decarads (10 heads + central unit) so the PVs listed above are doubled and differentiated by $(P).
Suggested $(P) = RADM:SYS0:100 or RADM:SYS0:200 for the two systems.

Introduction

The Decarad system was developed at JLab - it consists of 10 radiation detecting heads (geiger-mueller (GM) tubes) that connect (signal & HV cables) to a central box. That box has a board inside for changing the HV level and reading out the GM tube signals. The center box is connected to a digi port server via DB9 at box to RJ50 at the digi, which is then connected to the outside world via ethernet. The software uses the epics tool Stream with an asyn connection to the digi to send commands to the center box to read and write. The software is part of the RadiationMonitoring slac epics package.

Cable Connection

Notes on cable connection (from email 8/12/21)

RJ50 cable/breakout (digiport server side)                              DecaRad Chassis DB9
Pins                                                                                                  Pins
1 blue/wht  TX-                                                                             2
2 org/wht    RX-                                                                             8
3 org      NC
4 green/wht   NC
5 green TX+                                                                                    7
6 brown/wht  RX+                                                                         3
7 brown    SGND                                                                            6
8 blk/wht    NC
9 black         NC
10 blue        NC

Hardware Components

Using Decarad frame.pdf as the basic design

Thingsourcepart numberQty
for 2 frames
total
1.5x1.5" lite smooth 80-20

tecotechnology.com

prices from 8020.net

1515-LS-9610@72"
12@14"
12@24"

$7/foot comes to $700

$658.56

x

flange mount for caster to cross piece

I bought 100 3458 bolts

tecotechnology.com

2426
attach kit: 3458

24 flanges

4*24 bolts/nuts

24*$23

4*24*$1 = $650

x


caster
I bought 100 3110 bolts
teco tech2323
attach kit: 3110

24

4*24 bolts/nuts

24*$30

4*24*$0.3 = $750

x

x

5 hole tee plate for lower cross piece to vertical piece

I bought 190 3320 bolts

teco tech4340
attach kit: 3320

12
5*12 bolts

12*$10
5*12*$0.77 = $175

x

x

7 hole corner between end verticals and first upper horizteco tech4352
attach kit: 3320
4
7*4 bolts
4*$10.34
7*4*$0.77 = $65
x
6 hole T to connect upper horiz with verticals in the middleteco tech4310
attach kit: 3320
8
6*8 bolts

8*$10
6*8*$0.77 = $125

x
4 hole corner bracket to tie decarad heads to - only need two bolts perteco tech4301
attach kit: 3320
20
20*2
40*$5.12
40*2*$0.77 = $275
x
teco tech total (bob's total was $1k)


$2571.48 quote
wheel to hold 150' ethernet cable

2

75' retractable extension cord with 4 plugsamazon
22*$70 = $150
150' cat 6 ethernet cableBob says cablestogo
22*$110
cartmcmaster-carr2602T2422*$283.23 = $575
velcro zipties for decarads 15"long by 1/2" wide 5 for 9.73 greenmcmaster-carr6605K14408*9.73 = $78
velcro zipties for cables 11" long by 0.18" wide 10 for 8.39 bluemcmaster-carr6605K13505*$8.39 = $45
power strip with GFCImcmaster-carr7693K4922*$87.31 = $175
DB9 male connectorL-comC&P9M44*$1.10 = $5
DB9 pinsL-comCPM1044*$3.91 = $20
DB9 hoodL-comSDC9AG44*$3.81 = $16
L-com total


5+20+16=$41

wheel for 150' cable: https://www.amazon.com/Woods-E103-Extension-Holiday-Storage/dp/B000E8SWQM/ref=sr_1_1?keywords=woods+e103+extension+storage&qid=1639714062&sr=8-1 (no longer available)

cabestogo 150 ft cat 6: https://www.cablestogo.com/product/43170/150ft-cat6-snagless-solid-shielded-ethernet-network-patch-cable-blue

black & Decker retractable extension cord 75' with 4 plugs: https://www.amazon.com/dp/B091GSSDVV/ref=sspa_dk_detail_3?psc=1&pd_rd_i=B091GSSDVVp13NParams&smid=A38BOHDI5WNXSO&spLa=ZW5jcnlwdGVkUXVhbGlmaWVyPUEyODdYUlUwRkFJNjFJJmVuY3J5cHRlZElkPUEwODM4MDIxMVhZNUFRMEhaS0Y2RiZlbmNyeXB0ZWRBZElkPUEwMTkyNjcxMUtPMFJOUUo1RkxZTiZ3aWRnZXROYW1lPXNwX2RldGFpbDImYWN0aW9uPWNsaWNrUmVkaXJlY3QmZG9Ob3RMb2dDbGljaz10cnVl

Teco order was 110 3110, 110 3458, and 220 3320 bolts. 

Cable Lengths & Measurements

Ethernet cable length calculation: max(min(distance to nearest))+40' = 145' (CM33) Worst is 240' from CM midpoint (CM08)

120V 50' should be good (worst is 7' from end of H2)

Individual GM heads circumferences: 9.25" & 12.5" (w/o connectors)

Cables for heads: 7 short and 12 long (both types) plus 1 each unknown cuz in Mike's office (1/4/22)

Center box is 14" wide 14.75 incl connectors on front and 3.5" tall

Quotes & Receipts

Quote from Teco Tech:

 QUO14144.pdf

Receipt from Teco:

TECO_SO74053_01-04-2022-14-50-29.pdf

Original Files

Files from Bob and JLab folks (2020, 2021)
Com Cable Pinout.pdfDeca_stuff.tar.gzDecaRad_src.tar.gzdecarad_to_the_world.pdfdecarad_to_the_world.pptxDecRad System.docxiocnl22rfcry.tar.gzLND 714_DecaRad tube_datasheet.pdfSchematic.pdf80-20.xlsx80-20 Shopping Cart.pdfdecaraddocs.docxLND 714_DecaRad tube_datasheet.pdfDecRad System.docxDecarad frame.pdfdecarad_to_the_world.pdfhow long cables.pptx