intro: this doc will walk you through starting up the RGA, gathering data, and sending it off for qualification. It assumes you are using an MKS brand RGA.
highest pressure for RGA use: 7.6E-6 torr
[or, if you want to use faraday mode (but why would you want to do that?), max pressure = 5 x 10-4 Torr]
connection issues? see Computer Recognition of RGA
Gathering data:
double check:
- pressure is below 7e-10 torr
- valve to RGA is open
heck, double check all valves!
software settings:
filament on
accuracy 6
electron multiplier
(w/ the lowest option in the drop-down menu)
saving data
to prevent the computer falling asleep: open an image (any image) and put it on slideshow
*insert video of RGA startup*
Email data to Peter Stefan (CC Brady Robertson and Cynthia Melendrez) for official qualification judgement:
- The pressure that the cold cathode gauge reads as data collection stops.
- The data file. Found here:
- Cart number