Making note of which MPOD channels correspond to which detector (ePix10k small/Quad/2M, ePix100, others) and what the operating voltages are. 

ePix10ka small OR ePix100


Channel (cable s0)Channel (cable s0)VoltageCurrent
AVDD2002026
DVDD2012036
HV45200
PVDD204205*3.5

*Might vary based on temperature sensor reading. Typically ranges from 2-4 V (as long as temperature stays between 3-10 C). 


ePix10k2M

Table entries correspond to Q0/1/2/3


ChannelVoltageCurrent
**AVDD101/103/105/1076 all)~ 1.5
DVDD100/102/104/1066 (all)~ 4.3
HV0/1/2/3200 


** Turn all AVDD on first, wait 5-10 seconds, then turn DVDD on. (Kaz said it ensure 2M configures properly?) 




  • No labels