Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH2885 ± 125 39Graphs
FM2863 ± 11839Graphs
FL2849 ± 11740Graphs
AHL2895 ± 12439Graphs
AML2863 ± 11839Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH

5

269Graphs
FM25067Graphs
FL28245Graphs
AHL6198Graphs
AML34239Graphs


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH0
FM0
FL0
AHL9
AML7

Linearity (signal injection)

Thermal measurements (all thermometers)



  • No labels