Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH3052 ± 140 15Graphs
FM3053 ± 14015Graphs
FL3052 ± 14015Graphs
AHL3051 ± 14015Graphs
AML3051 ± 14015Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH675899Graphs
FM6135164Graphs
FL6135165Graphs
AHL6834Graphs
AML6135164Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH7
FM7
FL7
AHL7
AML7

Linearity (signal injection)

Thermal measurements (all thermometers)



  • No labels