Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH2995 ± 128 40Graphs
FM2960 ± 11844Graphs
FL2943 ± 11647Graphs
AHL2994 ± 12441Graphs
AML2958 ± 11844Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH5954Graphs
FM26711Graphs
FL28362Graphs
AHL6434Graphs
AML36398Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH0
FM0
FL48
AHL11
AML7

Linearity (signal injection)

Thermal measurements (all thermometers)