Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH3177 ± 237 5042Graphs
FM3139 ± 229 5062Graphs
FL3121 ± 224 5075Graphs
AHL3179 ± 233 5046Graphs
AML3140 ± 228 5065Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH6124546Graphs
FM3338139Graphs
FL2414233Graphs
AHL735416Graphs
AML3330343  Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH353
FM352
FL6016
AHL226
AML53

Configuration Settings

  

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History


ExperimentHutchShiftsDate Deployed
Commissioning