Performance Measures

Pedestal Summary

Method:  Dark environment, nominal operation, collect 1000 frames @120Hz.  Generate per-pixel mean and standard deviation over 1000 frames.  Pixels more than 3 sigma out of detector-wide mean are flagged bad.  Files archived as xxxx.

Map of pedestal non-conforming pixels for all gain modes here

Gain Mode

Mean Pedestal Value (ADU)

Number of NCP

Pixels Map, Histogram
FH3187 ± 236 5189Graphs
FM3145 ± 2275218Graphs
FL3126± 2215227Graphs
AHL3190 ± 2315196Graphs
AML3147 ± 2265218Graphs


Noise Summary

Map of noise non-conforming pixels for all gain modes here.  

Gain ModeVariation (ADU)Number of NCPPixels Map, Histogram
FH699508Graphs
FM3201231Graphs
FL2251297Graphs
AHL733245Graphs
AML3200612Graphs     


Gain and uniformity (photon source)

Map of gain non-conforming pixels for all gain modes here

Gain ModeNumber of NCP
FH3
FM2
FL6017
AHL258
AML62

Configuration Settings 

  

Linearity (signal injection)

Thermal measurements (all thermometers)

Experiment History


ExperimentHutchShiftsDate Deployed
Commissioning


  • No labels