Per-pixel gains from flat field measurement

DateDetectorGain modePhoton sourceOccupancy (photons/pixel)Per-pixel histogram fit HDF5Deployed gain fileComments
2016-02-16DsaCsPadhighIron K-alpha (6.4keV)0.015

/reg/g/psdm/detector/alignment/cspad/

calib-cxi-camera2-2016-02-05/calib/

CsPad::CalibV1/CxiDs2.0:Cspad.0/pixel_gain/

 run16_DsaCsPad_flat_orig.h5

cxi01516-r0016-2016-02-18-FeKalpha.dataSuccess rate was low ~0.75%. Filled in the rest with average gain from each asic.
2016-04-13DsaCsPadhighCu K-alpha (8.0keV)0.043

/reg/g/psdm/detector/alignment/cspad/

calib-cxi-camera2-2016-02-05/calib/

CsPad::CalibV1/CxiDs2.0:Cspad.0/pixel_gain/

 run6_DsaCsPad_flatIso.h5

run6_DsaCsPad_flatIso.data

Success rate was high ~0.99%, but there were some scattering in the centre, so filled that in with the average gain from each asic.

Note: CsPad has high and log gain modes. A factor of 7 is used to recalibrate the high gain to low gain, e.g. 28 ADU at high gain results in 4 ADU at low gain.

Calculation note: At 8keV, we expect around 30 ADUs per photon in high gain mode. At 6.4keV, we are seeing around 23 ADU per photon. currentADU/currentEnergy*8keV=23/6.4*8=28.75 ADU per photon which is close to 30 ADUs.

 

K-alpha table for Iron (6.4keV):

http://www.yourperiodictable.com/iron.php

K-alpha table for Copper (8.0keV):

http://www.yourperiodictable.com/copper.php