Note: these gain numbers are calculated using "flat field" images, where each pixel has a small number of photons (1-3). They may not be valid for high photon fluxes.
CsPad per-pixel gains from flat field measurement
Date | Detector | Gain mode | Photon source | Occupancy (photons/pixel) | Per-pixel histogram fit HDF5 | Experiment/Run | Deployed gain file | Comments |
---|---|---|---|---|---|---|---|---|
2016-02-16 | DsaCsPad | high | Iron K-alpha (6.4keV) | 0.015 | /reg/g/psdm/detector/alignment/cspad/ calib-cxi-camera2-2016-02-05/calib/ CsPad::CalibV1/CxiDs2.0:Cspad.0/pixel_gain/ run16_DsaCsPad_flat_orig.h5 | cxi01516/16 | cxi01516-r0016-2016-02-18-FeKalpha.data | Success rate was low ~0.75%. Filled in the rest with average gain from each asic. |
2016-04-13 | DsaCsPad | high | Cu K-alpha (8.0keV) | 0.043 | /reg/g/psdm/detector/alignment/cspad/ calib-cxi-camera2-2016-02-05/calib/ CsPad::CalibV1/CxiDs2.0:Cspad.0/pixel_gain/ run6_DsaCsPad_flatIso.h5 | cxi00516/6 | run6_DsaCsPad_flatIso.data | Success rate was high ~0.99%, but there were some scattering in the centre, so filled that in with the average gain from each asic. Experiment: cxi00516 run6 |
2018-04 | DsaCsPad | Cu | cxilr6716 | gain_prelim.npy | From Derek Mendez: Its not perfect but seems to improve the data (pretty important for the correlations experiment). Interestingly in this camera there is also a lone quad that is off. The initial goal was to measure an intensity-dependent gain correction (hence the different transmissions), but for now I just use the average gain cal from these data. One such average-gain map is stored as a 3D numpy array (same shape array returned by det.raw(event) ), this is in cxilt1417/scratch/gain/gain_prelim.npy* . the correction would be >> gain = np.load("gain_prelim.npy") >> assert ( np.all(gain!=0) ) >> data = det.calib(None) >> data /= gain |
Note: CsPad has high and low gain modes. A factor of 7 is used to recalibrate the high gain to low gain, e.g. 28 ADU at high gain results in 4 ADU at low gain.
Calculation note: At 8keV, we expect around 30 ADUs per photon in high gain mode. At 6.4keV, we are seeing around 23 ADU per photon. currentADU/currentEnergy*8keV=23/6.4*8=28.75 ADU per photon which is close to 30 ADUs.
K-alpha table for Iron (6.4keV):
http://www.yourperiodictable.com/iron.php
K-alpha table for Copper (8.0keV):
http://www.yourperiodictable.com/copper.php
Pnccd per-pixel gains from flat field measurement
Date | Detector | Gain mode | Photon source | Occupancy (photons/pixel) | Per-pixel histogram fit HDF5 | Experiment/Run | Deployed gain file | Comments |
---|---|---|---|---|---|---|---|---|
2016-04-15 | pnccdFront | 6 | Si K-alpha (1.74keV) | 0.006 | /reg/g/psdm/detector/alignment/pnccd/ amo06516-pnccd-2016-04-14/calib/ PNCCD::CalibV1/Camp.0:pnCCD.0/pixel_gain/ run10_pnccdFront_flatIso_clean.h5 | amo06516/10 | run10_pnccdFront_flatIso_clean.data | Success rate was high ~0.96%. Filled in the rest with average gain from each 128 pixel strips. |
2016-04-15 | pnccdFront | 5 | Si K-alpha (1.74keV) | 0.004 | /reg/g/psdm/detector/alignment/pnccd/ amo06516-pnccd-2016-04-14/calib/ PNCCD::CalibV1/Camp.0:pnCCD.0/pixel_gain/ run15_pnccdFront_flatIso_clean.h5 | amo06516/15 | run15_pnccdFront_flatIso_clean.data | Success rate was high ~0.96%. Filled in the rest with average gain from each 128 pixel strips. |
2016-04-15 | pnccdFront | 4 | Si K-alpha (1.74keV) | 0.003 | /reg/g/psdm/detector/alignment/pnccd/ amo06516-pnccd-2016-04-14/calib/ PNCCD::CalibV1/Camp.0:pnCCD.0/pixel_gain/ run19_pnccdFront_flatIso_clean.h5 | amo06516/19 | run19_pnccdFront_flatIso_clean.data | Success rate was high ~0.95%. Filled in the rest with average gain from each 128 pixel strips. |
Note: Pnccd has numbered gain modes.
Kaz says that PNCCD gives out ~1250ADU per 1keV photon in highest gain range.
K-alpha table for Silicon (1.74keV):
EPIX per-pixel gains
NOTE: In 2017 we have started using the epix id numbers with the new global calibration store. So this epix gain section is now deprecated. The global calibration store is at /reg/d/psdm/detector/calib/ and can be examined with the "dcs" command in psana.
Date | Photon Source | Carrier ID 0/1 | Digital Card ID 0/1 | Analog Card ID 0/1 | Experiment/Run | Gain File | Comment |
---|---|---|---|---|---|---|---|
2016-04-14 | Cu K-alpha (8 keV) | 996663297/3791650826 | 1232098304/953206283 | 2655595777/520093719 | xcs01116/95 | gainPixelCorr_passZero_r95_c1.txt | From Philip Hart |
2016-06-30 | Cu K-alpha | 996477441/2264924170 | 1794137088/403490827 | 2397900801/4076863512 | xcs06016/37 | gainPixelCorr_passZero_r37_c1.txt | From Philip Hart |
2016-06-30 | Cu K-alpha | 996579585/553648138 | 1232098304/1221641739 | 2650251521/3976200215 | xcs06016/37 | gainPixelCorr_passZero_r37_c2.txt | From Philip Hart |
2016-06-30 | Cu K-alpha | 996663297/3791650826 | 1232098304/953206283 | 2655595777/520093719 | xcs06016/37 | gainPixelCorr_passZero_r3335_c3.txt | From Philip Hart |
2016-07-01 | Ni fluorescence | 998779393/117440522 | 1794137088/403490827 | 2669921025/100663319 | xcs06016/52 | gainPixelCorr_passZero_r52_c2.txt | From Philip Hart |
2016-07-01 | Ni fluorescence | 996513537/2080374794 | 1794135040/940361739 | 2398406657/419430424 | xcs06016/52 | gainPixelCorr_passZero_r52_c3.txt | From Philip Hart |
2016-07-01 | Cu K-alpha | 996432897/3590324234 | 1232100352/1154532875 | 2654088449/33554455 | xcs06016/37 | gainPixelCorr_passZero_r37_c0.txt | From Philip Hart |
2016-10-21 | As above | xcsm9816/2-27,45-49 | gainPixelCorr_passZeroL_r2049_c2.txt | From Philip Hart | |||
2017-06-06 | Cu K-alpha | 997010433/3170893834 | 1794137088/403490827 | 2403244545/1224736792 | mfx01316/6,7 | gain3PixelCorr_passZero_r6007_c0.txt | From Philip Hart, fit statistics might be marginal so consider using the alternative non-fit results below |
As above | gainPixelCorr_passZero_r6007_c0.txt | From Philip Hart, uses mode instead of gaussian fit because of marginal statistics | |||||
2017-09-22 | Fe55 | 0996513537/2080374794 | 1794135040/0940361739 | 2398406657/0419430424 | detdaq17 9-15 | gainPixelCorr_passZeroL_r10121315_c0.npy | From Philiph Hart |
The best way to access epix10k ID numbers is with this command:
epix10ka_id exp=mfxc00318:run=13 epix10k2M
An uglier approach is to use the following script:
from psana import * import sys ds = DataSource(sys.argv[1]) evt = ds.events().next() cs = ds.env().configStore() cfg = cs.get(Epix.Config100aV2, Source(sys.argv[2])) print str(cfg.version()) # not ideal for epix10ka2M (and quads): currently the config DDL version (cfg.Version). ideally would be the firmware version (but didn't exist for early versions) print str(cfg.carrierId0())+'/'+str(cfg.carrierId1()) print str(cfg.digitalCardId0())+'/'+str(cfg.digitalCardId1()) print str(cfg.analogCardId0())+'/'+str(cfg.analogCardId1()) #a0 = cfg.asics(0) #a0.chipID()
Which can be run like this:
python epixSerialNumber.py exp=mecl8216:run=73 epix100a