Data

exp=cxif5315:run=147, 198 - dark

exp=cxif5315:run=162 - water sample

exp=cxif5315:run=169 - data sample

Detector: CxiDs2.0:Cspad.0

Calibration files

Use my local calibration store

[psana]
calib-dir = /reg/neh/home1/dubrovin/LCLS/rel-mengning/calib

Dark files were obtained and deployed using Calibration Management Tool

exp=cxif5315:run=147  dark average, rms, and average difference between runs 147-198:

Difference of average dark images between runs 147-198 is small, 0.3 ADU, difference rms=0.6 ADU, this proves that CxiDs2.0:Cspad.0 is stable during 50 runs.

 