Metrology data
Run 4 Ds1 metrology (added April 13)
Run 4 Ds1 average gain corrections
- /reg/neh/home/philiph/cxi/myana/gainAnalysis/hantkeGainCorrections_highCut_cleaned_relative.txt (using signal as a function of intensity)
- /reg/neh/home/philiph/cxi/myana/singlePhotonAnalysis/averagedCorrections_cleaned_relative.txt (using single photon peaks at low intensity)
- images at https://confluence.slac.stanford.edu/display/PCDS/CSPad+analysis+meeting%2C+10+August+11
- these should be sensible for any detector
Run 4 DSD metrology (pre-run) , image
- "There are 4 2x1 sensors that do not work. There is also a single 2x1 slot on one of the PCBs that has a bad connection. So the sensor works, but we can't operate it.
- The inner, beam-side 2x2s have tungsten shields, but only one other 2x2 does. Many of these modules were used during Run3 and so have some radiation damage and chipped sensors." - from Chris
Run 4 XPP metrology (pre-run) , image
- quad order in file is [0, 3, 1, 2] from looking at IZM modules - more info
- Mikhail's fits are at /reg/d/psdm/xpp/xpp47712/calib/CsPad::CalibV1/XppGon.0:Cspad.0/
- There are 9 types of files:
- center/0-end.data
- center_corr/0-end.data
- marg_gap_shift/0-end.data
- offset/0-end.data
- offset_corr/0-end.data
- quad_rotation/0-end.data
- quad_tilt/0-end.data
- rotation/0-end.data
- tilt/0-end.data
Run 5 DSD metrology,image, configuration
Run 5 XPP (solid plate) metrology for Mar 2012: spreadsheet, description
- Chris writes, "Here is the metrology for the camera on the new baseplate. All sensors locations are
relative to the lower, left corner of quadrant 3. Within each quadrant, I measured the sensor
corners starting form the lower left corner of each sensor and then moved clockwise around
the chip. - With the new mechanical package doing this by quadrants doesn't make sense. In the future, it
might be better to simply organize the metrology by vertical columns. - The 2x2 module numbering is also attached. It's the same as last time except two modules were
swapped."
- The metrology was done upside down, so q2->q0 etc and x->177457-x, y->185867-y
- photograph
Run 5 XPP low gain calibrations
- see 14 Mar 12 CSpad meeting page for comments
- fullFrameAbsGain_xppcom10_lowGain_noFN.txt
- fullFrameAbsGain_xppcom10_lowGain_fn.fits
Run 6 DS2 aka DSD CSPAD V1.2 metrology
Run 7 XPP V1.2 metrology
Run 7 MEC 140k metrology
- metrology , related images - under OGP , 1-5 label mapping to module , order of survey points
Run 7 DS1 v1.2 metrology
2013-10-09-CSPAD-XPP-Metrology.xlsx
2013-12-10-Quad-MEC-Metrology.xlsx
2013-12-12-CSPAD-CXI-DSD-Metrology.xlsx
2013-12-20-CSPAD-CXI-DS1-Metrology.xlsx
2014-03-19-CSPAD-CXI-DS1-Metrology.xlsx
2 May 2014 Files and comment from Gabriel about recent metrology work:
- The DS1 detector was rebuilt on March 12-17 2014.
- Thus for the DS1:
- the Milathianaki experiment and later should use the 2014-03-19 metrology
- the Yachandra experiment and earlier should use the previous 2013-12-20 metrology
- The DS2 should use the 2013-12-12 metrology
2014-05-15-CSPAD-CXI-DS1-Metrology.xlsx
2014-05-15-CSPAD-CXI-DS2-Metrology.xlsx
27 Oct 2014: Files and comment from Gabriel
2014-08-27-CXI-DS2-Metrology.xlsx
2014-09-24-CXI-DS1-Metrology.xlsx
Attached, the most recent CXI metrology files. These replace the May 2014 versions and should be used for all experiments after September 24, 2014.
Changes in CXI-DS1:
- Quad 3 (zero based counting) reworked; metrology updated in the attached file: "DS1-Metrology-Sept-24-2014"
- Q3 Z values measured properly
- Mikhail, please double check Q3 values for consistency
- The other 3 quads (Q0, Q1, Q2) are unchanged (as in the previous metrology: May 15 2014).
- Please note that this camera has been renamed. Possible new name: "CXI Camera 1"+ version number. Jason, could you please confirm the exact name and version.
Changes in CXI-DS2:
- Quad 1 (zero based counting) reworked; metrology updated in the attached file: "DS2-Metrology-Aug-27-2014"
- Q1 Z values: measured incorrectly (OGP bug?), please discard.
- Mikhail, could you please check Q1 values for consistency?
- The other 3 quads (Q0, Q2, Q3) are unchanged (as in the previous metrology: May 14, 2014).
- Please note that this camera has been renamed. Possible new name: "CXI Camera 2"+ version number. Jason, could you please confirm the exact name and version.
As always, these excel metrology files are NOT FOR DISTRIBUTION to avoid user confusion. Only Mikhail's curated version should be distributed and used.