You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 5 Next »

Recommended Runs for Data Analysis

Go get the from the runs database

First calibration runs for experts
  + CPT_beamtest FM109 TEM1 700000392 PASSED INT 01.66.2
  + CPT_beamtest FM101 TEM3 700000431 PASSED INT 01.66.2
  + CPT_beamtest FM119 TEM2 700000418 PASSED INT 01.66.2
  + calibDAC FM109 TEM1 700000400 FAILED INT 1.66.2
  + calibDAC FM119 TEM2 700000438 FAILED INT 1.66.2
  + calibDAC FM101 TEM3 700000445 FAILED INT 1.66.2
  + calibGen CU 15MeV 700000458 PASSED INT 1.66.2
  + muTrig   CU 15MeV 7200s 700000462 PASSED INT 1.66.2
  + calf_gain_p01_1.28 FM109 TEM1 700000486
  + calf_gain_p01_1.28 FM119 TEM2 700000485
  + calf_gain_p01_1.28 FM101 TEM3 700000487

Muon calibration runs BT-13 (4rng, nonZS, muon Gain, CAL_LO 15MeV)
   10 runs 700000276 to 700000286
   10 runs 700000490 to 700000500

Synchronized runs (check tags in rcReport and txt files in run directory) :
  + 700000475 : 10k events, external trigger on scintillators.
  + 700000502 : external trigger on scintillators, new ancillary DAQ

Calibrations

Expected value for most important registers in DAC Counts

CAL LAC : 15-20 
CAL FLE : 15-20 
CAL FHE : 15-20 
CAL ULD : 15-20 
CAL FHE : 15-20 

TKR DAC : 15-20 
 

Timing

The values shown below are based on the LAT Alignment and we will use these as default for CU tests. Note that this is for the flight configuration B2 that became LAT701 config. TEM diagnostics are not enabled and they will change these values.

 
 

Below is a timing diagram that explains how knobs are set to ensure charge is collected at the peak for all 3 detectors. Note that CAL beling the slowest its TREQ Delay is set to 0, while ACD being the fastest detector has more delays added.

 
 

Finally there are some plots from trigger jitter from Martin Kocian!LAT trigger jitter.gif!
 
 

  • No labels