Requirements to meet:


RequirementsMeasured resultnote
Mode of operationAHL (functional)Verified

AML (functional)Verified

FH (functional)Verified

FM (functional)Verified

FL (functional)Verified
Pixel size100x100 um2Verified
Range

Autoranging: >40000 keV

(transitions 400 keV/1200 keV selectable)




FH >400 keV


FM >1,200 keV


FL > 40,000 keV

Noise r.m.s.

<~400eV (110 e- rms)



Note that the specifications must meet the requirements at a readout rate that exceeds 5102 Hz or 4644 Hz.