Using Rohde & Schwarz HMP2020 power supply

The SMA-F input labeled CONTROL is limited from 0V to 10V.  This voltage is used to set the gain of the pulse.  If you want a pulse of ~10V, then set V_CONTROL = 1V


Steps
1Hit Menu
2Down arrow to "Over Voltage Protection (OVP)
3Right arrow to select OVP
4Use arrows and/or scroll wheal to set for CH1: 10V← Max voltage CONTROL Port can take



→ Connect CH1 to SMA-F _TRIGGER port & terminate with 50 Ohm

SettingValueNote(s).............................................
ModePulse
Impedance50 Ohm
Period1ms← In photo used frequency, but easier with period
HighLevel3.3V
LowLevel0mV
Pulse Width990us← Want a 10us pulse low (needs to be high most the time for proper operation)
Rise Edge6ns
Delay0.0ns