SA-261-101-34-C00_TYP_USE.pdf

SA-261-101-34-C00_TYP_USE.vsdx

  1. Limit CONTROL Power Supply
    1. Using Rohde & Schwarz HMP2020 power supply

      Make sure output is off

      The SMA-F input labeled CONTROL is limited from 0V to 10V.  This voltage is used to set the gain of the pulse.  If you want a pulse of ~10V, then set V_CONTROL = 1V


      Steps
      1Hit Menu
      2Down arrow to "Over Voltage Protection (OVP)
      3Right arrow to select OVP
      4Use arrows and/or scroll wheal to set for CH1: 10V← Max voltage CONTROL Port can take



  2. Turn on Power Supply 
    1. Turn on this supply

      This supply is used to provide +/-15V to the pulser box: 

      CH[V]
      1+15
      2-15
      3N/A


  3. Turn On COMP (Source Meter) to offset pulse +10V (DC bias of substrate)
    1. Using rear outputs as they are limiting the Keithley 2410 SourceMeter is 1.1kV.  The black-box limiter in the back limits the current to 5mA and voltage to 240V.

      SettingValueNote(s)..............
      FRONT/REARREAR
      LIMIT CURRENT105uAReverse biasing substrate, as long as we don't lock up the substrate by turning on the parasitic thyristor (PNPN), current consumption is very lower
      SOURCE+10V

      Make sure to plug cable from rear to SMA-F COMP port on front of pulser-box (see green arrow in photo)

  4.  Setup and turn on _TRIGGER 
    1. → Connect CH1 to SMA-F _TRIGGER port & terminate with 50 Ohm

      SettingValueNote(s).............................................
      ModePulse
      Impedance50 Ohm
      Period1ms← In photo used frequency, but easier with period
      HighLevel3.3V
      LowLevel0mV
      Pulse Width990us← Want a 10us pulse low (needs to be high most the time for proper operation)
      Rise Edge6ns
      Delay0.0ns
  5. AC coupled OUT of Pulser box to oscilloscope (1MOhm load) and DET_BIAS of camera 

  6. Increase the voltage on CONTROL Power Supply (MAX 10V)
    1. Output voltage step size = 10mV
    2. OPTIONAL: If a finer sweep is required (1mV) a ATT 10X has been provided (it is unidirectional

      ← OPTIONAL!!!

    3. As you increase the voltage you'll you see an ~10X increase in the pulser-ouput height voltage

      1. CONTROL [V]Photo

        1

        4

  7. If pulser is working correctly will see LED light come on




  • No labels