Population of the calibration store with constants is a significant issue for instrument scientists and users.
Code
Essential code is implemented under the Detector package in modules app/deploy_constants and src/UtilsDeployConstants.py
Command
In psana environment on any psana node (with access to data) use command
deploy_constants -h
(ana-4.0.42) [dubrovin@psanagpu105:~/LCLS/con-py2]$ deploy_constants -h
usage:
deploy_constants -e <experiment> -d <detector> -r <run-number> -C <calib-type> [-D] [-L <logging-mode>] [...]
REGULAR COMMAND EXAMPLE:
deploy_constants -e xpptut15 -r 260 -d XcsEndstation.0:Epix100a.1 -C gain -D -c ./calib -D --runrange 10-20
Input parameters to select dataset: # DataSource argument:
-e xpptut15 -r 260 # exp=xpptut15:run=260
-e xpptut15 -r 260 -s :smd # exp=xpptut15:run=260:smd
-e xpptut15 -r 260 -s :dir=path/xtc::stream=0-79 # exp=xpptut15:run=260:dir=path/xtc::stream=0-79
-x /cds/data/drpsrcf/mfx/mfxlv4920/xtc/<file-name.xtc> # /cds/data/drpsrcf/mfx/mfxlv4920/xtc/<file-name.xtc>
TEST COMMANDS:
Ex.1 for epix100a : deploy_constants -e xpptut15 -r 260 -d XcsEndstation.0:Epix100a.1 -C gain -D -c ./calib
Ex.2 for Cspad CXI : deploy_constants -e xpptut15 -r 380 -d CxiDs1.0:Cspad.0 -C gain -D -c ./calib
Ex.3 for Cspad CXI : deploy_constants -e xpptut15 -r 380 -d CxiDs2.0:Cspad.0 -C gain -D -c ./calib
Ex.4 for Cspad XPP : deploy_constants -e xpptut15 -r 320 -d XppGon.0:Cspad.0 -C gain -D -c ./calib
Ex.5 for cspad2x2 : deploy_constants -e xpptut15 -r 460 -d MecTargetChamber.0:Cspad2x2.3 -C gain -D -c ./calib
Ex.6 for pnCCD : deploy_constants -e xpptut15 -r 450 -d Camp.0:pnCCD.1 -C gain -D -c ./calib
Help: deploy_constants -h
Deployment of calibration parameters
optional arguments:
-h, --help show this help message and exit
-e EXP, --exp EXP experiment name, default = None
-d DET, --det DET detector name, default = None
-r RUN, --run RUN run number of the existing dataset to get run time-
stamp, detector full name etc., default = None
-t TSTAMP, --tstamp TSTAMP
non-default time stamp (YYYYmmddHHMMSS) to search for
constants in repo. By default run time is used,
default = None
-s DSSUFFIX, --dssuffix DSSUFFIX
dataset name suffix beyond
exp=<expname>:run=<runnum><dssuffix>, for example
":smd:dir=<path>/xtc/", default = None
-x DSNAME, --dsname DSNAME
overrides exp=<expname>:run=<runnum><dssuffix> if
specified, e.g. path to *.xtc file, default = None
-o DIRREPO, --dirrepo DIRREPO
non-default repository of calibration results, default
= /reg/g/psdm/detector/calib/constants/
-c DIRCALIB, --dircalib DIRCALIB
non-default deployment calib directory, different from
standard one, default = None
-C CTYPE, --ctype CTYPE
calibration type [gain, common_mode, etc.], default =
gain
-l LOGLEV, --loglev LOGLEV
logging mode, one of INFO, CRITICAL, WARN, WARNING,
ERROR, DEBUG, NOTSET, default = INFO
--dirmode DIRMODE mode for all mkdir, default = 511
--filemode FILEMODE mode for all saved files, default = 438
-D, --deploy deploy constants to the calib dir, default = False
-R RUNRANGE, --runrange RUNRANGE
validity run range for output calibration file,
default = 0-end
Repository
Repository structure is defined for flexibility to account for detector dependent time dependent and default constants
Repository:
repodir = /reg/g/psdm/detector/calib/constants/
Log files:
<repodir>/logs/ - directory for log files
<repodir>/logs/<YEAR>/<time-stamp>_log_deploy_constants_<uid>.txt
ex: <repodir>/logs/2022/2022-06-09T121926_log_deploy_constants_dubrovin.txt
Detector type directory:
<repodir>/<dettype> - detector type directory for dettypes: cspad, cspad2x2, epix100a, pnccd
<repodir>/<dettype>/.aliases.txt - dictionary between detector id/name and short alias
Directory for default files:
<repodir>/<dettype>/default/
<dettype>_default_<ctype>.txt
ex: epix100a_default_gain.txt
Specific detector directory:
<repodir>/<dettype>/<detector-id-or-name>/
Calibration type subdirectory:
<repodir>/<dettype>/<detector-id-or-name>/<calib-type>/
<dettype>_<alias>_<YYYYmmddHHMMSS>_<exp-name>_r<run-number>_<ctype>.txt
ex: epix100a_0001_20100101000000_xpptut15_r0260_gain.txt
Content
Currently a few files with constants are supplied for epix100a gain, common_mode and geometry
/reg/g/psdm/detector/calib/constants/epix100a/default/epix100a_default_common_mode.txt, epix100a_default_gain.txt, epix100a_default_geometry.txt
/reg/g/psdm/detector/calib/constants/epix100a/3925999616-0996579585-0553648138-1232098304-1221641739-2650251521-3976200215/gain/epix100a_0001_20100101000000_xpptut15_r0260_gain.txt
References