(Identification Details and Hardware Changelog
Detector ID: 996579585
Nominal Operating Voltages and Currents
Ch. | Line | V | I |
---|---|---|---|
Performance Measures
Experiment=detdaq18, Run=267 (Sep 3, 2020)
Summary
For Pedestals, method: Dark environment, nominal operation, collect 1000 frames @120Hz. Generate per-pixel mean and standard deviation over 1000 frames. Pixels more than 3 sigma out of detector-wide mean are flagged bad. Files archived as xxxx.
exp:run (mm/dd/yy) | Mean Pedestal Value (ADU) | Number of Ped NCP | Pedestal Graphs | Noise (ADU) | Number of Noise NCP | Noise Graphs | |
---|---|---|---|---|---|---|---|
detdaq19, run 263 (11/11/21) | Histogram/Pixel Map Pedestal NCP Map | Histogram/Pixel Map Pedestal NCP Map | |||||
detdaq18, run 267 (9/3/20) | 3908 ± 224 | 26 | 19 | 3332 |
Thermal measurements (all thermometers)
Configuration Settings
Experiment History
Run 19
Experiment | Hutch | Shifts | Date Deployed (mm/dd/yy) |
---|---|---|---|
LW64 | XPP | 9/22/21 |