Camera Model | Pulse Polarity | Pulse Delay | Pulse Width | comments | ||
---|---|---|---|---|---|---|
Adimec Opal 1000 | Pos | 0.0008 | 0.0002 | Pulse width affects exposure time. 0.0001 sec is too short. | ||
Adimec Opal 4000 | Pos | 0.0008 | 0.0002 | Maximum trigger rate is 30Hz. | ||
Pulnix TM-6740CL | Neg | 0.0006 | 0.0005 | Pulse width affects exposure time. | ||
LBL FCCD | Pos |
|
| Pulse width does not affect exposure time. 60 Hz strongly recommended. | ||
Cspad | Neg | 5.59092e-4 | 1e-5 | Ryan says will change to Pos polarity during shutdown | ||
Cspad2x2 | Pos | 8.65762e-4 | 1e-5 | Critical that the "Run" trigger coming from controls evr have polarity | 1e-5 is fine. In seconds, this delay is 2.982e-4. | The new firmware in the Cspad2x2 has moved the integration window |
Cspad2x2 | Pos | 5e-4 | 1e-5 | Run trigger the same, 5e-4, or 59524 | ||
Febo (normal) | Pos | 7.78e-4 | 1e-6 | For normal mode, trigger width not critical | ||
Febo (Average) | Pos | 4.41e-4 | 1e-6 | For averaging mode, trigger width not critical | ||
IPIMB | Neg | 0.000651824 | 5e-06 |
|
...