Recommended Runs for immediate Data Analysis

Go get the from the runs database

...

Synchronized runs (check tags in rcReport and txt files in run directory) :
  + 700000475 : 10k events, external trigger on scintillators.
  + 700000502 : external trigger on scintillators, new ancillary DAQ

Calibrations

Expected value for most important registers in DAC Counts

CAL LAC : 15-20 
CAL FLE : 15-20 
CAL FHE : 15-20 
CAL ULD : 15-20 
CAL FHE : 15-20 

TKR DAC : 15-20 
 

Timing

The values shown below are based on the LAT Alignment and we will use these as default for CU tests. Note that this is for the flight configuration B2 that became LAT701 config. TEM diagnostics are not enabled and they will change these values.

 Image Removed
 

Below is a timing diagram that explains how knobs are set to ensure charge is collected at the peak for all 3 detectors. Note that CAL beling the slowest its TREQ Delay is set to 0, while ACD being the fastest detector has more delays added.

 Image Removed
 

CAL Energy Calibrations

Includes results from calibrations prior to CERN 

Analysis of muon data taking in the CPT runs is included here 

Also dedicated muon runs results are found here

CAL CPT Charge injection runs

Includes data analysis results on all charge injection runs, table generation etc... 

Baseline Data Taking

The baseline definition for the registers used for data taking is found here.Finally there are some plots from trigger jitter from Martin Kocian!LAT trigger jitter.gif!