Updated default run for intensity scans:

Results:

Link to analysis plan:

Beamtime run notes (Bhavna):

https://docs.google.com/document/d/1eYhR0q_LjSd4DGc9OnYuBl2aFFucNxmGOBq_E0NMT5g/edit

Analysis plan (Bhavna):

https://docs.google.com/spreadsheets/d/10m2vQLeCrlIA53khhRLaTRqm4YGobh0RGqGaE3A_WL8/edit

Primary finding for smalls: the detector is marginally timed in, which is causing the low values after switching. The delay, ACQToAsicR0delay, should be shifted from 18223 to 17223.

Google Drive folder for presentations with results etc.:

https://drive.google.com/drive/u/1/folders/1oolbkUqPAMD76qWZYiGQvVKIQPCqUfBS

Initial simulations of effects for getting the timing wrong. (Conny)